脉搏波形态学研究及其在子痫前期的应用 被引量:3

Morphology of Photoplethysmography and Its Application in Preeclampsia
下载PDF
导出
摘要 子痫前期是妊娠20周后出现的特有疾病,会导致血液系统、消化系统、神经系统等的异常改变,严重危害孕产妇和围生儿的健康.针对实际临床中存在的子痫前期检测方法复杂、不便于日常监护的问题,提出了基于光电容积脉搏波提取相关特征参数的方法.根据脉搏波下降支的形态特征及生理意义,提取出了新的特征参数——光电容积斜率指数(PSI),通过逐段分析下降支斜率来体现外周血管阻力、血管壁弹性等生理信息的细节变化.基于临床采集的50例孕妇(23名为子痫前期患者,27名为正常孕妇)的光电容积脉搏波数据,对PSI指数区分子痫前期患者与正常孕妇的有效性进行了验证.结果表明,真阳性20例,假阳性1例,真阴性26例,假阴性3例,灵敏性和特异性分别达到87.0%和96.3%,准确性达到92.0%.PSI指数为子痫前期的无创诊断研究提供了理论依据和新的研究方法,对于实现子痫前期的早期筛查、保障孕妇和胎儿的健康有着重要意义. Preeclampsia(PE)is a disorder that may develop after 20 weeks of pregnancy,which could increase the risk of poor outcomes for both the mothers and the fetuses.Specifically,it could lead to abnormal changes in the hematologic,digestive,and nervous systems.As the diagnostic method of PE used in clinical practice is quite complex, there is an urge demand that a more convenient method should be proposed.In this thesis,a novel parameter extracted from the photoplethysmography (PPG) was proposed to improve the situation above.The new characteristic parameter,photoplethysmography slope index(PSI),was conducted from both the morphological features and the physiological significances of PPG.By analyzing the slope of a segmented descending curve,the diversity of the PPG was evaluated,which contained detailed physiological information including perivascular resistance and vascular wall elasticity.The ability of PSI to distinguish PE from normotensive pregnancies was also assessed in this thesis.The PSI was contrasted,which were calculated from the PPG data of 50 pregnancies (27 without and 23 with PE).The results revealed that there were 20 true positives,1 false positive,26 true negatives,and 3 false negatives.The receiver operating characteristic analysis indicated that the PSI could predict PE with a sensitivity of 87.0%,a specificity of 96.3%,and an accuracy of 92.0%.The PSI provided a new noninvasive method for PE diagnosis. It was suggested that PSI might be of significance for early PE screening and for the health of the mothers and fetuses.
作者 陈婉琳 江锋 陈新忠 封英 缪家骏 焦翠翠 陈杭 Chen Wanlin;Jiang Feng;Chen Xinzhong;Feng Ying;Miao Jiajun;Jiao Cuicui;Chen Hang(College of Biomedical Engineering and Instrument Science,Zhejiang University,Hangzhou 310027,China;Key Laboratory of Biomedical Engineering of Ministry of Education,Zhejiang University,Hangzhou 310027,China;Zhejiang Provincial Key Laboratory of Cardio-Cerebral Vascular Detection Technology and MedicinalEffectiveness Appraisal,Zhejiang University,Hangzhou 310027,China;Women’s Hospital,School of Medicine,Zhejiang University,Hangzhou 310027,China)
出处 《天津大学学报(自然科学与工程技术版)》 EI CSCD 北大核心 2019年第8期857-861,共5页 Journal of Tianjin University:Science and Technology
基金 国家自然科学基金资助项目(81870868)~~
关键词 光电容积脉搏波 子痫前期 光电容积斜率指数 形态学 photoplethysmography(PPG) preeclampsia(PE) photoplethysmography slope index(PSI) morphology
  • 相关文献

参考文献1

二级参考文献33

  • 1Ahonen, J., Jokela, R., Uutela, K., Huiku, M., 2007. Surgical stress index reflects surgical stress in gynaecological laparoscopic day-case surgery. British Journal of Anaesthesia, 98(4):456-461. [doi:10.1093/bja/aem035].
  • 2Alastruey, J., Parker, K.H., Peiro, J., Sherwin, S.J., 2009. Analysing the pattern of pulse waves in arterial networks: a time-domain study. Journal of Engineering Mathematics, 64(4):331-351. [doi:10.1007/s10665-009-9275-1].
  • 3Allen, J., 2007. Photoplethysmography and its application in clinical physiological measurement. Physiological Measurement, 28(3):R1-R39. [doi:10.108810967-3334/28/3/R01].
  • 4Bhattacharya, J., Kanjilal, RP., Muralidhar, V., 2001. Analysis and characterization of photo-plethysmographic signal. IEEE Transactions on Biomedical Engineering, 48(1): 5-11. [doi: 10.1109/10.900243].
  • 5Eyal, S., Oz, O., Eliash, S., Wasserman, G., Akselrod, S., 1997. The diastolic decay constant in spontaneously hypertensive rats versus WKY rats as an indicator for vasomotor control. Journal of the Autonomic Nervous System, 64(1):24-32. [doi:10.1016/s0165-1838(97)00012-X].
  • 6Fogliardi, R., Burattini, R., Shroff, S.G., Campbell, K.B., 1996. Fit to diastolic arterial pressure by third-order lumped model yields unreliable estimates of arterial compliance. Medical Engineering & Physics, 18(3):225-233. [doi:10. 1016/1350-4533(95)00042-9].
  • 7Gnudi, G., 1998. New closed-form expressions for the estimation of arterial Windkessel compliance. Computers in Biology and Medicine, 28(3):207-223. [doi:10.1016/ 90010-4825(98)00008-0].
  • 8Haffty, B.G., O'Hare, N.E., Singh, J.B., Spodick, D.H., 1983. Noninvasive tracking of peripheral resistance by ear densitography. Chest, 83 (5):771-775. [doi:10.1378/chest. 83.5.771].
  • 9Hashimoto, J., Chonan, K., Aoki, Y., Nishimura, T., Ohkubo, T., Hozawa, A., Suzuki, M., Matsubara, M., Michimata, M., Araki, T., et al., 2002. Pulse wave velocity and the second derivative of the finger photoplethysmogram in treated hypertensive patients: their relationship and associating factors. Journal of Hypertension, 20(12): 2415-2422.
  • 10Huiku, M., Uutela, K., van Gils, M., Korhonen, I., Kymalainen, M., Merilainen, P., Paloheimo, M., Rantanen, M., Takala, P., Viertio-Oja, H., et al., 2007. Assessment of surgical stress during general anaesthesia. British Journal of Anaesthesia, 98(4):447-455. [doi:10. 1093/bja/aem004].

同被引文献18

引证文献3

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部