基于EPID的鼻咽癌容积旋转调强放疗计划的三维剂量验证的研究 被引量:15

Three-dimensional dose verification study of volumetric modulated arc therapy radiotherapy for nasopharyngeal carcinoma based on EPID
下载PDF
导出
摘要 目的利用基于电子射野影像装置(EPID)的EDose 5.0系统验证鼻咽癌(NPC)容积旋转调强放疗(VMAT)计划的三维剂量分布,探讨该系统用于鼻咽癌VMAT计划三维剂量验证的可行性。方法分别使用基于EPID的EDose 5.0、Delta^4和ArcCHECK三维剂量验证系统对随机选取的14例鼻咽癌VMAT计划进行验证测量,并在不同Gamma(γ)分析标准下,比较不同三维验证系统的验证通过率。结果当选取3%/3 mm标准时,基于EPID的EDose 5.0、Delta^4和ArcCHECK的γ平均通过率均大于95%,分别为(97.09±1.40)%、(96.24±1.52)%和(97.40±1.68)%,均满足计划验证要求;EDose 5.0与Delta^4和ArcCHECK验证通过率两两比较表明,当评价标准为3%/3 mm、3%/2 mm和2%/2 mm时,EDose 5.0与Delta^4和ArcCHECK的测量验证结果间差异无统计学意义(P>0.05)。结论基于EPID的EDose 5.0系统可以用于鼻咽癌VMAT计划的三维剂量验证。 Objective To verify the 3D dose distribution of the volumetric modulated arc therapy(VMAT) plan for nasopharyngeal carcinoma(NPC) by using the EDose 5.0 system based on the electronic portal imaging device(EPID),and to analyze the feasibility of the system for three-dimensional dose verification of nasopharyngeal carcinoma VMAT plan. Methods A total of 14 nasopharyngeal carcinoma VMAT plans were selected randomly.These plans were measured and verified by EPID-based EDose 5.0,Delta^4 and ArcCHECK three-dimensional dose verification systems,and the passing rates of different 3D verification systems were compared under different gamma(γ) analysis criteria. Results When the 3%/3 mm standard was selected,the average passing rate of γ based on EPID-based EDose 5.0,Delta^4 and ArcCHECK was greater than 95%. The specific measuring results of three-dimensional dose verification were(97.09±1.40)%,(96.24±1.52)% and (97.40±1.68)% respectively,which all met the planning verification requirements;When EDose 5.0 was compared with the verification rates of Delta^4 and ArcCHECK respectively under the criteria of 3%/3 mm,3%/2 mm and 2%/2 mm( P >0.05),the result showed that there was no significant difference between the measurement results of EDose 5.0 and of Delta^4 and ArcCHECK. Conclusion The EPID-based EDose 5.0 system can be used for three-dimensional dose verification of the VMAT plan for nasopharyngeal carcinoma.
作者 张朋 闫冰 薛旭东 刘磊 吴爱林 吴爱东 Zhang Peng;Yan Bing;Xue Xudong(College of Life Sciences,Anhui Medical University,Hefei 230032;Dept of Radiation Oncology,The Affiliated Provincial Hospital of Anhui Medical University,Hefei 230001)
出处 《安徽医科大学学报》 CAS 北大核心 2019年第5期758-761,共4页 Acta Universitatis Medicinalis Anhui
基金 国家自然科学基金(编号:11805198) 安徽省自然科学基金青年项目(编号:1808085QH281)
关键词 鼻咽癌 容积旋转调强放疗 质量控制 电子射野影像装置 EDose nasopharyngeal carcinoma volumetric modulated arc therapy quality control electronic portal imaging device EDose
  • 相关文献

参考文献6

二级参考文献48

  • 1McDermott LN, Louwe R J, Sonke JJ, et al. Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device. Med Phys,2004 ,31:285-295.
  • 2Louwe RJ, McDermott LN, Sonke JJ, et al. The long-term stability of amorphous silicon fiat panel imaging devices for dosimetry purposes. Med Phys,2004,31:2989-2995.
  • 3Ben'y SL, Polvorosa CS, Wuu CS. A field size specific backscatter correction algorithm for accurate EPID dosimetry. Med Phys, 2010,37:2425-2434.
  • 4Winkler P , Georg D. An intcreomparison of 11 amorphous silicon EPIDs of the same type: implications for portal dosimetry. Phys Med Biol,2006,51:4189-4200.
  • 5McDermott LN, Nijsten SMJJG, Sonke JJ, et al. Comparison of ghosting effects for three commercial a-Si EPIDs. Med Phys,2006, 33:2448-2451.
  • 6Rowshanfarzad P, Sabet M, O'Connor DJ, et al. Reduction of the effect of non-uniform backscatter from an E-type support arm of a Varian a-Si EPID used for dosimetry. Phys Med Biol,2010,55: 6617-6632.
  • 7Siebers JV, Kim JO, Ko L, et al. Monte Carlo computation of dosimetric amorphous silicon electronic portal images. Med Phys, 2004,31:2135-2146.
  • 8Parent L, Scco J, Evans PM, et al. Monte Carlo modelling of a-Si EPID response:the effect of spectral variations with field size and position. Med Phys, 2006, 33:4527-4540.
  • 9Parent L, Fielding AL, Dance EPID calibration for dosimetric DR, et al. Amorphous silicon applications: comparison of a method based on Monte Carlo prediction of response with existing techniques. Phys Med Biol,2007,52:3351-3368.
  • 10Wang S, Gardner JK, Gordon JJ, et al. Monte Carlo-based adaptive EPID dose kernel accounting for different field size responses of imagers. Med Phys,2009,36:3582-3595.

共引文献49

同被引文献141

引证文献15

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部