摘要
介绍了奥林巴斯CV-260图像处理装置的内部结构特点及其具体功能,详细分析了开机黑屏与开机反复重启2例故障的维修过程。指出了维修此类故障需要了解装置的结构和原理、仔细观察故障现象、推导故障发生原因,为医疗同行维修类似故障提供了参考。
The internal structural characteristics and specific functions of Olympus CV-260 image processing device were introduced. The maintenance of blank screen and repeated restarting after startup was analyzed deeply, and it's suggested the technician understand the structure and principle of the device and explore the failures and their causes carefully.
作者
黄世安
于巍
HUANG Shi-an;YU Wei(Department of Medical Engineering,NO. 980 Hospital of Joint Logistics Support Force,Shijiazhuang 050082,China)
出处
《医疗卫生装备》
CAS
2019年第6期105-107,共3页
Chinese Medical Equipment Journal