期刊文献+

JMS SP-500微量注射泵故障分析及维修 被引量:3

Analysis and Repair of JMS SP-500 Micro Injection Pump
下载PDF
导出
摘要 本文通过分析医院在用、且具备一定市场占有率的JMS品牌SP-500型号微量注射泵的声光报警、注射器无法识别或识别后闪烁报警的故障原因,提出切实可行的维修建议,为微量注射泵的维修提供了借鉴经验。同时,本文基于医院设备管理及至生物医学工程行业日益重视的质量控制工作理念,从实际工作经验出发,提出维修后按规程进行质控检测的建议,从而保障临床科室使用正常使用且质量合格、计量精准的微量注射泵,保障医疗质量,避免不良事件。 By analyzing the reasons of the sound-light alarm,the unrecognized syringe or the flashing alarm after identification of the JMS sp-500 micro-injection pump,which was in use in the hospital and had a certain market share,the feasible maintenance suggestions were put forward to provide reference experience for the maintenance of micro-injection pump.At the same time, based on the concept of quality control that hospital equipment management and biomedical engineering industry pay more and more attention to,based on the practical work experience,the suggestion of quality control test according to the rules after maintenance was put forward in this paper,so as to ensure the use of normal,qualified and accurate micro-injection pump in clinical departments,guarantee the medical quality and avoid a dverse events.
作者 麦达成 肖翔 王永胜 MAI Dacheng;XIAO Xiang;WANG Yongsheng(Department of Medical Equipment,Guangzhou Panyu Central Hospital,Guangzhou Guangdong 511400,China)
出处 《中国医疗设备》 2019年第7期170-172,175,共4页 China Medical Devices
关键词 微量注射泵 声光报警故障 质量控制 可疑不良事件 micro injection pump sound and light alarm failure quality control suspected adverse events
  • 相关文献

参考文献15

二级参考文献88

共引文献176

同被引文献40

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部