频域光学相干断层扫描测量正常人中央角膜厚度的可重复性研究 被引量:1

Reproducibility of the measurement of central corneal thickness of normal persons with spectral-domain optical coherence tomography
下载PDF
导出
摘要 目的评估频域光学相干断层扫描(SD-OCT)测量正常人中央角膜厚度(CCT)的可重复性。方法随机选取正常人50例共50眼,应用SD-OCT测量CCT,扫描深度为2.0 mm,光源波长为840 nm,轴向分辨率为6μm,横向分辨率为20μm,用系统自带标尺测量角膜顶点前表面到后表面的垂直距离,记录CCT,评估重复性。结果同一观察者测量方法的相关系数(ICC)为0.9558(95%CI:0.925~0.978),变异系数(CV)为2.5%。不同观察者测量方法的ICC为0.9335(95%CI:0.912~0.966),CV为4.8%。观察者1第1次测量的CTT值为(538.53±32.37)μm,第2次为(543.5.25±32.85)μm,差异无统计学意义(P>0.05)。观察者2测量的CCT为(524.53±28.37)μm,与观察者1两次测量的结果比较差异均有统计学意义(P<0.05或0.01)。结论 SD-OCT测量正常人CCT的可重复性高;应用SD-OCT追踪CCT的变化时,应该由同一检测者进行测量。 Objective To evaluate the reproducibility of the measurement of central corneal thickness (CCT) of normal persons with spectral-domain optical coherence tomography (SD-OCT). Methods A total of 50 eyes of 50 normal persons were selected, and their CCTs were measured using SD-OCT. The scanning depth was 2.0 mm, the light source wavelength was 840 nm, the axial resolution was 6 μm, and the lateral resolution was 20 μm. The vertical distance from the front surface to the rear surface of the corneal vertex was measured with the ruler of the system, and CCT was recorded to evaluate repeatability. Results The ICC of intraobserver measurement method was 0.9558 (95% CI:0.925-0.978), and the CV of intraobserver measurement method was 2.5%. The ICC of interobserver measurement method was 0.9335 (95% CI:0.912- 0.966), and the CV of intraobserver measurement method was 4.8%. The CTT of the first measurement by Observer 1 was (538.53±32.37)μm, and that of the second measurement by Observer 1 was (543.5.25±32.85)μm, and there was no statistical difference (P>0.05). The CCT measured by Observer 2 was (524.53±28.37)μm, which was statistically different from the results of two measurements by Observer 1 (P<0.05 or 0.01). Conclusion The reproducibility of CCT measurements using SD-OCT is high. The measurement shall be done by the same observer when SD-OCT is used to track the change in CCT.
作者 郑彪 彭红娟 黄婷 柯毅 林桂英 ZHENG Biao;PENG Hong-juan;HUANG Ting;KE Yi;LIN Gui-yin(Department of Ophthalmology, the Affiliated Hospital of Guangdong Medical University, Zhanjiang 524001, China)
出处 《广东医科大学学报》 2019年第4期437-439,共3页 Journal of Guangdong Medical University
关键词 频域光学相干断层扫描 中央角膜厚度 可重复性 spectral-domain optical coherence tomography central corneal thickness reproducibility
  • 相关文献

参考文献4

二级参考文献31

  • 1Hasan Ali Bayhan,Seray Aslan Bayhan,Iz.zet Can.Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter[J].International Journal of Ophthalmology(English edition),2014,7(2):302-308. 被引量:9
  • 2Kim JS, Ishikawa H, Sung KR, Xu J, Wollstein G, Bflonick RA, et al. Retinal nerve fiber layer thickness measurement reproduc- ibility improved with spectral domain optical coherence tomo- graphy[ J]. Br J Ophthalmo1,2009,93 ( 8 ) : 1057-1063.
  • 3Schuman JS. Spectral domain optical coherence tomography for glaucoma( an AOS thesis ) [ J ]. Trans Am Ophthalmol Soc, 2008,106 ( 3 ) :426-458.
  • 4Carpineto P, Ciancalglini M, Zuppardi E, Falconio G, Doronzo E, Mastropasqua L, et al. Reliability of nerve fiber layer thickness mesurements using optical coherence tomography in normal and glaucomatous eyes [ J ]. Ophthatmology, 2003,110 ( 1 ) : 190-195.
  • 5Shrout PE. Measurement reliability and agreement in psychiatry [J]. Stat Methods Med Res ,1998 , 7 ( 3 ) :301-317.
  • 6Paunescu LA, Schuman JS, Price LL, Stark PC, Beaton S, lsmka- wa H, et al. Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using Stratus OCT[ J]. Invest Ophthalmol Vis Sci ,2004,45 (6) : 1716-1724.
  • 7Budenz DL, Fredette MJ, Feuer WJ, Anderson DR. Reproducibli- ty of peripapillary retinal nerve fiber thickness measurements with stratus OCT in glaucomatous eye [ J ]. Ophthalmology, 2008,115 (4) :661-666.
  • 8Chan A, Duker JS, Ko TH, Fujimoto JG, Schuman JS. Normal macular thickness measurements in healthy eyes using Stratus optical coherence tomography [ J ]. Arch Ophthalmol, 2006,124 (2) :193-198.
  • 9Hsu SY, Tung IC, Sheu MM, Tsai RK. Reproducibity of peripapil- lary of retinal nerve fiber layer and macular retinal thickness measurements using optical coherence tomography[ J]. Kaohsi- ung J Med Sc/,2006,22(9) :447-451.
  • 10Sehi M, Guaqueta DC, Feuer WJ, Greenefield DS. A comparison of structural measurements using Stratus optical coherence tomography instruments [ J]. Glaucoma, 2007,16 ( 3 ) : 287 -292.

共引文献28

同被引文献2

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部