摘要
目的应用计算流体力学(computational fluid dynamic, CFD)技术,针对一款自行设计的导管泵进行数值研究,预测其水力性能和血栓形成风险。方法选用Grigioni与Danny Bluestein分别提出的血栓预测模型,并结合CFD技术计算血小板运动过程受到的剪切力与曝光时间,从而获得血小板活性状态(platelet activation state, PAS),实现对导管泵血栓的预测。结果流量为4L/min、转速为10000r/min时,导管泵扬程为14.763kPa,可满足左心室辅助的要求;Grigioni与Danny Bluestein预测模型下的PAS分别为6.35×10^-6和7.68×10^-4,数值较小,血栓形成性较低。结论探索基于数值方法的血栓预测的可行性,并预测导管泵的水力性能和血栓形成可能性,为后续的设计优化提供依据。
Objective To investigate a self-designed catheter pump by using computational fluid dynamics(CFD) method, so as to predict its hydraulic performance and risk of thrombosis formation. Methods The thrombosis prediction models proposed by Grigioni and Danny Bluestein were used. The shear stress and exposure time during platelet motion were calculated by CFD method, and parameters of platelet activation state(PAS) were obtained for prediction of thrombogenic performance. Results At the flow rate of 4 L/min and rotating speed of 10 000 r/min, the differential pressure of the pump reached 14.763 kPa and the hydraulic performance was proved to fit the requirement of left ventricular assist device. The PAS values of Grigioni model and Danny Bluestein model were 6.35×10^-6 and 7.68×10^-4,respectively, both at a very low level, indicating a low possibility of thrombus formation. Conclusions This study investigated the feasibility of thrombosis prediction based on simulation method and the predicted hydraulic performance and thrombosis will provide references for further design optimization.
作者
王芳群
吴怡
吉敬华
贺万堑
吴义荣
WANG Fangqun;WU Yi;JI Jinghua;HE Wanqian;WU Yirong(School of Electrical and Information Engineering,Jiangsu University,Zhenjiang 212013,Jiangsu,China)
出处
《医用生物力学》
EI
CAS
CSCD
北大核心
2019年第3期283-288,共6页
Journal of Medical Biomechanics
基金
国家自然科学基金项目(51677082,51477068)
关键词
导管泵
血栓
计算流体力学
血小板活性状态
catheter pump
thrombosis
computational fluid dynamics(CFD)
platelet activation state(PAS)