摘要 由于一些老旧仪器设备性能和技术指标还基本良好,仍然具有非常大的使用价值,但毕竟一些部件使用率极高,损坏程度大,即便是更换配件原厂价格也非常高,而且费时费工极易损坏原电路板.本文主要介绍利用微型继电器解决JSM-35C型扫描电子显微镜中放大系统琴键开关因长时间高频率使用产生接触不良和电阻导致图像不能稳定的问题,从而保证了原仪器设备电路的完整,既省时省工,又节约了资金. Some of the old instruments with basically goodperformance and technical indicators, can still be in service for a long time. However, the more frequently used parts are easier to worn out. It is not always practical to replace the OEM parts, due to the financial cost, time consuming and the potential further damage to the motherboard. This paper focused on the use of micro-relays to resolve the unstable image issue from JSM-35 scanning electron microscope, due to the resistor performance degradation caused by the high usage of its key switch. This resolution maintained the integrity of the original motherboard, while keeping the overall cost and time effort minimum.
作者 李炳太 LI Bing-tai(Electronic Microscope Room, Shanxi Medical University, TaiyuanShanxi 030001,China)
出处 《电子显微学报》 CAS CSCD 北大核心 2019年第4期414-415,共2页 Journal of Chinese Electron Microscopy Society
关键词 微型继电器 图像不稳 琴键开关 micro-relay unstable image key switch
  • 相关文献

参考文献2

二级参考文献10

  • 1M. Sutoh et al.A Micro-Fluxgate Magnetic Sensing Element Using Electrodeposited NiFeIn Films[].Technical Digest of The ~(th) Sensor Symposium in Japan.1995
  • 2Hashimoto Etsu,Tanaka Hedenao et al.Thermally Controlled Magnetization Actuator forMicrorelays[].IEICE Transactions on Electronics.1997
  • 3Ren H,Gerhard E.Design and Fabrication of A Current-pulse-excited Bistable MagneticMicroactuator[].Sensors and Actuators.1997
  • 4Zhou Shifang,Sun Xiqing,Carr William.N.MicroVariable Inductor Chip Using MEMS Relays[].International Conference on Solid-State Sensorand Actuators Proc IEEE.1997
  • 5Schiele Ignaz,Huber Joerg et al.Micromechanical Relay with Electronic Actuation[].InternationalConference on Solid-State Sensorand Actuators Proc IEEE.1997
  • 6Simon Jonathan,Saffer Scott,KimChang-Jin.Micromechanical Relay with Eletronstatic Actuation[].International Conference on Solid-State Sensorand Actuators Proc IEEE.1997
  • 7Ignaz Schicel et al.Micromechanical Relay with Electrostatic Actuation[].Tranducer’.1997
  • 8Desheng LI,Kyoseko Chun,Tetsuhiko Iizuka and Hiroyuki Fujita.Two Dimensional Micro MagneticSensor Array[].The Transactions of The Institute of Electrical Engineers of Japan.1998
  • 9E.Fillin et al.A New Basic Technology For Magnetic Micro-Actuators[].IEEE Transactions on Electron Devices.1998
  • 10Grigorishin I.L,Mukurov N.I,Efremov G.I.Series of Microcommutation Electrostatic Devices withA Wide Range of Operating Voltages[].IVMC IEEE NJ USA.1996

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部