期刊文献+

微量进样器校准方法 被引量:5

Calibration method of micro-sampling syringe
下载PDF
导出
摘要 建立了微量进样器校准方法.采用衡量法纯水和水银两种不同介质和分光光度法研究微量进样器的校准,通过对不同方法进行比较分析,得出最佳校准方法是采用纯水介质的衡量法.校准条件:控制环境温度在(20±5)℃,保证水温与室温温差不超过±2℃,相对湿度控制在30%~60%,采取防蒸发措施.采用该操作方法能够完全适用于微量进样器的校准. The calibration method of micro sampling syringe was established. Calibration of micro sampling syringe was studied by measuring pure water and mercury two different media and spectrophotometry. By comparing and analysising of different methods,it was concluded that the best calibration method was to use pure water medium measurement method. Calibration conditions: the ambient temperature was controlled at (20±5)℃, the temperature difference between water temperature and room temperature was not exceed ±2℃, and the relative humidity was controlled at 30%–60%, and take anti-evaporation measures. The operation method can be fully applied to the calibration of the micro sampling syringe.
作者 周秉直 臧鹏 田晓明 周方 ZHOU Bingzhi;ZANG Peng;TIAN Xiaoming;ZHOU Fang(Shaanxi Institute of Metrology Science , Xian 710068, China)
出处 《化学分析计量》 CAS 2019年第5期104-108,共5页 Chemical Analysis And Meterage
基金 国家质量监督检验检疫总局科技计划项目(2015QK200)
关键词 微量进样器 校准方法 衡量法 分光光度法 micro sampling syringe calibration method measurement spectrophotometry
  • 相关文献

参考文献5

二级参考文献9

  • 1McMillan N, O'Neill, M, Smith S, et al. Reliable TLDA-microvolume UV spectroscopy with applications in chemistry and biosciences for microlitre analysis and rapid pipette calibration I J]. Journal of Optics A: Pure and Applied Optics,2009, 11(5):1 -10.
  • 2Batista E, Arias R, Wang J T. Final report on BIPM/ CIPM key comparison CCM. FF-K4. 2. 2011: Volume comparison at 100 ixL--Calibration of micropipettes ( piston pipettes ) [ J ]. Metrologia, 2013,50 : 07003.
  • 3Batista E, Pinto L, Filipe E, et al. Calibration of micropipettes : Test methods and uncertainty analysis [ J ]. Measurement: Journal of the International Measurement Confederation,2007, 40 ( 3 ) : 338 - 342.
  • 4Aibe V Y. Water mass measurement m parts for volume standard calibration in tlie key comparison [ C ]// FLOMEKO, 14th International Flow Measurement Conference, Sandton, South Africa, 2007 : 145 - 148.
  • 5Madee T, Amarouehe S. Comparison of mieropipettes calibration by accredited laboratories [ C ]//3rd International Metrology Conference 2010, CAFMET 2010, 2010, Marrakech, Morocco,2010:337 - 343.
  • 6杨禹哲,廖鹏.农药残毒快速测定仪校准用标准物质及方法的研究[J].计量学报,2010,31(5A):78-79.
  • 7Batista E, Richard P. The selection of water property formulae for volume and flow calibration [ J ]. Metrologia, 2007, 44 (6) : 453 - 463.
  • 8Tanaka M, Girard G, Davis R, et al. Recommended table for the density of water between 0C and 40C based on recent experimental reports [ J ]. Metrologia, 2001,38: 301 -309.
  • 9李亚萍,张广军,李庆波.空间双光路红外CO_2气体传感器及其测量模型[J].光学精密工程,2009,17(1):14-19. 被引量:20

共引文献16

同被引文献23

引证文献5

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部