期刊文献+

山西永济地震台BBVS-60地震计标定异常原因分析

Cause Analysis of Calibration Anomaly of BBVS-60 Seismometer at Yongji Seismological Station in Shanxi Province
下载PDF
导出
摘要 针对山西永济地震台BBVS-60地震计在脉冲标定及正弦标定中出现的异常进行分析,排查原因,找出地震计和数据采集器在安装、使用及运输过程中存在的问题,进行解决,并对今后的仪器运维工作提出合理的改进建议。 In view of the abnormalities of BBVS-60 seismometer in pulse calibration and sinusoidal calibration at Yongji seismological station in Shanxi province, this paper analyses the causes,finds out the problems existing in the installation, use and transportation of seismometer and data acquisition device, and solves them, and puts forward reasonable suggestions for improving the operation and maintenance of seismometer in the future.
作者 邰垣东 张红秀 高志伟 TAI Yuan-dong;ZHANG Hong-xiu;GAO Zhi-wei(Linfen Center Seismologic Station of Earthquake Administration of Shanxi Province, Linfen, Shanxi 041000, China;National Continental Rift Valley Dynamics Observatory of Taiyuan, Taiyuan, Shanxi 030025, China)
出处 《山西地震》 2019年第3期54-56,共3页 Earthquake Research in Shanxi
关键词 BBVS-60地震计 标定异常 故障分析 BBVS-60 seismometer Calibration anomaly Fault analysis