针对山西永济地震台BBVS-60地震计在脉冲标定及正弦标定中出现的异常进行分析,排查原因,找出地震计和数据采集器在安装、使用及运输过程中存在的问题,进行解决,并对今后的仪器运维工作提出合理的改进建议。
In view of the abnormalities of BBVS-60 seismometer in pulse calibration and sinusoidal calibration at Yongji seismological station in Shanxi province, this paper analyses the causes,finds out the problems existing in the installation, use and transportation of seismometer and data acquisition device, and solves them, and puts forward reasonable suggestions for improving the operation and maintenance of seismometer in the future.