国产首台医用加速器图像引导系统(EPID)的质量保证与质量控制 被引量:6

Quality Assurance and Quality Control of the First Domestic Medical Accelerator Image Guidance System: EPID
下载PDF
导出
摘要 目的研究国产加速器图像引导系统(EPID)的质量保证内容方案和质量控制测试方法。方法使用验证模体和工具来验证图像引导系统(EPID)的安全碰撞连锁、系统运动、几何精度、配准精度、成像图像质量等。结果 EPID的安全连锁和系统运动正常,几何精度、配准精度、成像图像质量误差≤±2 mm或优于厂家推荐标准和要求。结论通过选择EPID合适的质量保证和质量控制,可以保证其在临床放疗中发挥重要的辅助摆位验证位置功能。 Objective To study the quality assurance content scheme and quality control test method of domestic accelerator image guidance system: EPID. Methods The secure collision chain, system motion, geometric accuracy, registration accuracy, and image quality of the image guidance system: EPID were verified using verification phantoms and tools. Results The safety chain and system movement of EPID were normal. The errors of geometric accuracy, registration accuracy, and imaging quality were ≤±2 mm or better than the manufacturer’s recommended standards and requirements. Conclusion By selecting the appropriate quality assurance and quality control of EPID, it can play an important role in the auxiliary position verification position in clinical radiotherapy.
作者 程长海 闫清华 郭洪涛 李定杰 宋迎新 CHENG Changhai;YAN Qinghua;GUO Hongtao;LI Dingjie;SONG Yingxin(Department of Radiotherapy,Zhoukou Central Hospital,Zhoukou Henan 466000,China;Department of Radiotherapy,The Tumor Hospital Affiliated to Zhengzhou University,Zhengzhou Henan 450000,China;Shandong Shinva Medical Devices Co.,Ltd.,Zibo Shandong 255086,China)
出处 《中国医疗设备》 2019年第10期23-26,共4页 China Medical Devices
基金 2016年国家高端医疗设备应用示范项目(2016MHD11004001100) 国家重点研发计划数字诊疗装备研发重点专项(2016YFC0106700)
关键词 直线加速器 图像引导系统 质量保证 质量控制 linear accelerator image guidance system quality assurance quality control
  • 相关文献

参考文献11

二级参考文献55

  • 1马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧.窄带野胶片检查在多叶准直器质量保证中作用探讨[J].中华放射肿瘤学杂志,2004,13(2):123-127. 被引量:10
  • 2沈捷,张福泉,邱杰,胡克,张明杰.应用EPID对放射治疗摆位误差的研究[J].中华放射医学与防护杂志,2006,26(3):308-308. 被引量:39
  • 3鞠忠建,王运来,马林,徐寿平,戴相昆,王连元.用二维电离室矩阵验证多叶准直器叶片到位精度[J].中华放射肿瘤学杂志,2006,15(4):335-338. 被引量:18
  • 4Michael G. Herman, James M. Baiter, David A. Jaffray, el at. Clinical use of electronic portal imaging:report of AAPM radiation therapy committee task group 58 [J]. Med. Phys., 2001,28(5):712-737.
  • 5J.W. Motz , M.Danos. Image information content and patient exposure [J]. Med. Phys., 1978,5(1): 8-22.
  • 6Carla D.Bradford, Walter W. Peppler, Jacob M. Waidelich. Use of a slit camera for MTF measurements [J]. Med, Phys., 1999,26 (11): 2286-2294.
  • 7Amit Sawant, Larry antonuk, Youcef EI-Mohri. Silt design for efficient and accurate MTF measurement at megavohage x-ray energies [J]. Med. Phys., 2007,34(5):1535-1545.
  • 8I. A. Cunningham, B. K. Reid. Signal and noise in modulation transfer function determinations using the slit, wire, and edge techniques[J]. Med. Phys., 1992,19(4):1037-1044.
  • 9Ehsan Samel, Egbert Buhr, Paul Granfors, et al. Comparison of edge analysis techniques for the determination of the MTF of digital radiographic systems [J]. Phys. Med. Biol., 2005,50:3613-3625.
  • 10Jean-Pierre Moy. Signal-to-noise ratio and spatial resolution in x-ray electronic imagers: Is the MTF a relevant parameter [J]. Med. Phys., 2000, 27(1): 86-93.

共引文献108

同被引文献44

引证文献6

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部