DSA 机房内 X 射线辐射剂量率监测结果分析 被引量:2

The Monitoring Results Analysis of X - ray Radiation Dose Rate in DSA Equipment Room
下载PDF
导出
摘要 通过对DSA不同工况下、不同位置X射线辐射空气吸收剂量率监测分析,揭示剂量分布较高区域和位置,并就监测结果结合瞬时测量值的仪器响应时间修正计算,对DSA机房及周围环境进行分析,为DSA机房内手术医护人员剂量估算、辐射防护评价与辐射防护管理提供参考。 By monitoring and analyzing the X ray dose rate of air absorbed in different position with different DSA conditions,the position monitored higher radiation dose were found.The correction calculation was conducted by combining the real-time monitoring result with the response time of the machine.This work could be used to estimate the radiation dose of medical staffs who work in DSA room,Furthermore,this work would provide a reference to estimate and assess the radiation dose and strengthen the radiation protection.
作者 杨文栋 武荣国 蒋文静 YANG Wen-dong;WU Rong-guo;JIANG Wen-jing(Yunnan Radiation Monitoring Station,Kunming Yunnan 650032,China)
出处 《环境科学导刊》 2019年第6期96-98,共3页 Environmental Science Survey
关键词 数字减影血管造影(DSA) X射线 辐射剂量监测 DSA X-ray radiation dose rate monitoring
  • 相关文献

参考文献3

二级参考文献17

  • 1张引,苗金萍,王巧云,邢茹,李毓玲.包头市介入放射治疗工作人员的受照射剂量[J].中华放射医学与防护杂志,2005,25(2):166-167. 被引量:5
  • 2高晓菊.从介入放射学中的辐射防护看皮肤剂量实时监测的必要性[J].中国医学装备,2004,1(1):25-26. 被引量:1
  • 3卫生部令第46号,放射诊疗管理规定[S].2006.
  • 4中华人民共和国国家质量监督检验检疫总局,GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2002
  • 5凡生,张林波,曹澎,等.DSA技术在血管造影中的应用[J].医疗卫生装备,2010,31(4):408-412.
  • 6Owings MF, Kozak LJ. Ambulatory and inpatient procedures in the United States, 1996[J]. Vital and health statistics. Series 13, Data from the National Health Survey, 1998 (139) : 1.
  • 7Koenig TR, Mettler FA, Wagner LK. Skin Injuries from Fluoroscop- ically Guided Procednres: Part 2. Review of 73 Cases and Recom- mendations for Minimizing Dose Delivered to Patient [ J ]. AJR, 2001, 177:13-20.
  • 8Koenig TR, Wolff D, Mettler FA, et al. Skin injuries from fluoro- scopically guided procedures: Part 1. Characteristics of Radiation In- jury[J]. AJR, 2001, 177:3-11.
  • 9Stranden E, Widmark A, Sekse T. Assessing doses to interventional radiologists using a personal dosimeter worn over a protective apron [J]. Acta Radiologica, 2008, 49(4) : 415 -418.
  • 10Sanchez R, Vano E, Femandez JM, et al. Staff Radiation Doses in a Real- Time Display Inside the Angiography Room[ J]. Cardiovasc Intervent Radiol, 2010, 33 : 1210 - 1214.

共引文献36

同被引文献14

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部