摘要 目的 :对日本OLYMPUSAU40 0全自动生化分析仪进行应用评价。方法 :采用两种不同浓度质控血清和患者及其健康人的血清分别就ALT、TG、Urea和TBiL四项指标对该仪器进行精密度对照试验、准确度试验、线性范围及预期偏倚测试。结果 :该仪器所测ALT、TG、Urea批内及总CV <3 % ;TBiL批内 <3 % ,总CV <10 % ;三种试剂比较r在 0 .984~ 0 .9992间 ,ALT、TG、TBiL、Urea相对偏倚分别小于 3 .5 %、3 .5 %、9.5 %、8.6%。结论 :参照NCCLS文件 ,该仪器主要性能指标评价符合文件要求 ,OLYMPUSAU40 0全自动生化分析仪具有良好的精密度、准确度、线性范围、试剂开放特点 。 Objective: To use and evaluate Olympus AU400 automated chemistry analyzer which is made in Japan. Methods: The precision,accuracy,the linear range and deviation index of ALT,TG,Urea and TBil in the serums of two kinds of control quality,patients and health people were detected. Results: The withinrun cv and total withinrun cv of ALT,TG,Urea are both less than 3%,The withinrun cv of TBil is less than 3%,total withinrun cv is less than 10%,the correlation coefficients of three kinds of reagents is 0.984 ~ 0.9992 ,the deviation index is less than 3.5% , 3.5% , 9.5% and 8.6% .Conclusion: The precision,accuracy,recovery rate and linear range of Olympus AU400 are satisfaction.It can suit for clinical biochemistry laboratory and scientific research.
作者 张洪波
出处 《江汉大学学报(医学版)》 2002年第3期41-43,共3页 Journal of Wuhan Professional Medical College
关键词 OLYMPUS AU400 全自动化分析仪 应用 精密度 预期偏倚 automatic biochemistry analyzer precision deviation index
  • 相关文献

参考文献3

  • 1NCCLS (EPS-T2). Evaluation of Precision of Clinical Chemistry Devices-Second Edition; Tentative guideline[s].
  • 2NCCLS (EP9-A). National Committee for Clinical Laboratory Standards. Methold of Comparision and Bais Estimation Using Patient Sample; Approved Guideline[S].
  • 3NCCLS (EP6-P). National Committee for Clinical Laboratory Standards. Evaluation of the Linearity of Quantitative Analytical Methods; Proposed Guidline[ S].

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部