摘要
Tansmission grating spectrometers are extensively used to measure absthlute X-ray spectra in a photon-energy range below 1000 eV. The transmission grating,as its dispersive element, must be calibrated to obtain its danaction efficiencies. Calibrations of absolute distraction efficiencies of the tlansmission grating at photon energyof 844eV have been carried out on Beliing Synchrotron mediation Facility. With theaid of grating model, all of the grating structure paraxneters have been determinedand the absolute dchaction efficiencies in a photon-energy range below 2000eV havealso been calculated and discussed.
Tansmission grating spectrometers are extensively used to measure absthlute X-ray spectra in a photon-energy range below 1000 eV. The transmission grating,as its dispersive element, must be calibrated to obtain its danaction efficiencies. Calibrations of absolute distraction efficiencies of the tlansmission grating at photon energyof 844eV have been carried out on Beliing Synchrotron mediation Facility. With theaid of grating model, all of the grating structure paraxneters have been determinedand the absolute dchaction efficiencies in a photon-energy range below 2000eV havealso been calculated and discussed.
关键词
同步加器辐射
衍射效率
软X射线
Synchrotron radiation
Transmission grating
Diffraction efficiency