摘要 目的对某一特定猕猴设计一款高填充系数、高信噪比的多通道接收线圈,并通过仿真验证其可行性。材料与方法通过电磁场及高速电路仿真软件,对线圈的形状、位置、线圈间的去耦、线圈的调谐匹配等进行仿真,进而得出最优的线圈设计。结果利用LC去耦网络及重叠面积去耦法完成了四通道线圈去耦仿真,并由此确定线圈的形状位置以及电路的各项参数。结论本文通过仿真验证了对某一特定猕猴设计一款定制多通道接收线圈的可行性,并得到一系列设计参数,为后续的线圈制作以及实际线圈的电路调试等简化设计流程,降低研发成本。 Objective:In this study,a multi-channel receive-only coil with high filling coefficient and SNR(signal-to-noise ratio)was designed for a given macaque head model.Furthermore,the feasibility of the design was examined using simulation test.Materials and Methods:We use electromagnetic field and high-speed circuit simulation software to examine the tuning and the decoupling effect of the multi-channel coils and to optimize our design.Results:We calculated the decoupling effects of four-channel coil with LC-decoupling network and overlapping method,and further confirmed the final design and documented the electronic characters of the receive-only coil.Conclusions:This paper illustrated the feasibility to design a multi-channel receiver coil for a given macaque with simulation test.In addition,the parameters of our design could provide a protocol for other coil designs,which could reduce the research cost.
作者 傅方杰 徐俊成 赵超 姚守权 蒋瑜 FU Fangjie;XU Juncheng;ZHAO Chao;YAO Shouquan;JIANG Yu(Shanghai Key Laboratory of Magnetic Resonance,School of Physics and Material Science,East China Normal University,Shanghai 200062,China)
出处 《磁共振成像》 CAS 2019年第8期600-603,共4页 Chinese Journal of Magnetic Resonance Imaging
基金 国家自然科学基金项目(编号:21874045)~~
关键词 磁共振成像 猕猴 多通道接收线圈 去耦 仿真 magnetic resonance imaging macaque multi-channel receive-only coil decoupling simulation
  • 相关文献

参考文献3

二级参考文献17

  • 1董宝玮,梁萍.肿瘤热消融治疗:现状和展望[J].中华医学杂志,2006,86(12):793-796. 被引量:24
  • 2CLINE H E, HYNYNEN K, HARDY C J, et al. MR temperature mapping of focused ultrasound surgery [J]. Magnetic Resonance in Medicine, 1994, 31(6), 628-636.
  • 3CLINE H E, HYNYNEN K, HARDY C J, et al. MR temperature mapping of focused ultrasound surgery [J]. Magnetic Resonance in Medicine, 1994. 31(6), 628-636.
  • 4RIEKE V, PAULY K B. MR Thermometry [J]. Journal of Magnetic Resonance Imaging, 2008, 27(2): 376-390.
  • 5GERMAIN D, VAHALA E, EHNHOLM G J, et al. MR temperature measurement in liver tissue at 0. 23 T with a steady-state free precession sequence [J]. Magnetic Resonance in Medicine, 2002, 47(5), 940-947.
  • 6KOPELMAN D, INBAR Y, HANANNEL A, et al. Magnetic resonance-guided focused ultrasound surgery (MRgFUS) :Ablation of liver tissue in a porcine model [J]. Eur J Radiol, 2006, 59(2):157-162.
  • 7ROEMER P B, EDELSTEIN W A, HAYES C E, et al. The NMR phased array [J]. Magnetic Resonance in Medicine, 1990, 16(2): 192-225.
  • 8LEE R F, GIAQUINTO R O, HARDY C J. Coupling and decoupling theory and its application to the MRI phased array [J]. Magnetic Resonance in Medicine, 2002, 48(1): 203- 213.
  • 9RIEKE V, ROSS A B, NAU W H, et al. MRl-temperature mapping during ultrasound prostate ablation using fat for phase estimation [J]. Conf Proc IEEE Eng Med Biol Soc, 2004, 4: 2500-2502.
  • 10Daniel Papoti,Cecil Chern‐Chyi Yen,Julie B. Mackel,Hellmut Merkle,Afonso C. Silva.??An embedded four‐channel receive‐only RF coil array for fMRI experiments of the somatosensory pathway in conscious awake marmosets(J)NMR Biomed. . 2013 (11)

共引文献4

同被引文献7

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部