期刊文献+

Beckman-CoulterLH750全自动血细胞分析仪性能验证

Performance verification of Beckman-Coulter LH750 automatic hematology analyzer
下载PDF
导出
摘要 目的:对BeCkman-CoulterLH750全自动血细胞分析仪进行性能验证,以保障检验结果的准确可靠。方法:参照中华人民共和国卫生行业标准(WS/T406-2012),并对性能指标的本底计数、准确度、精密度(重复精密度、日间精密度)、携带污染率、线性分析进行检测。结果:本底计数、准确度、精密度、携带污染率、线性分析均在要求范围内。结论:Beckman-CoulterLH750全自动血细胞分析仪性能良好检测结果准确可靠,可用于临床血常规分析检查。 Objective:To evaluate the performance of Beckman—Coulter LH750 automatic hematology analyzer in order to ensure the accuracy and reliability of testing results.Methods;Background count,Carrying pollution rate,precision,linearity,trueness were evaluated according to the verification method of PRC heath industry standard(WS/T406—2012).Results:The background county carrying pollution rate,precision,linearity,trueness were within the permissible range of quality requirements.Conclusion:Beckman-Coulter LH750 automatic hematology analyzer has good performance,and its detection results are accurate and reliable can used in clinical laboratories.
作者 苑高翔 朱保忠 YUAN Gao-xiang;ZHU Bao-zhong(Equipment division,Fourth people's hospital of Taiyuan,Shanxi Taiyuan 030053)
出处 《医学检验与临床》 2019年第9期26-29,共4页 Medical Laboratory Science and Clinics
关键词 全自动血细胞分析仪 性能评价 Automatic hematology analyzer Performance verification Evaluation
  • 相关文献

参考文献3

二级参考文献19

  • 1廖益,王冬梅,金平,李晋蜀.Sysmex XT-2000i高端血细胞分析仪性能概况综合评价[J].实用医学杂志,2006,22(21):2528-2530. 被引量:4
  • 2杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127.
  • 3杨琛懋,马璟.关于GLP实验室仪器验证的探讨[J].毒理学杂志,2007,21(4):285-285. 被引量:10
  • 4陈永红,刘正明,张文敬,胡浩.法国ABX PENTRA60五分类血球分析仪传感超时原因分析及排除方法[J].西北国防医学杂志,2007,28(5):387-387. 被引量:5
  • 5冯仁丰.临床实验室建设与管理[M].上海:上海科学技术出版社,1991:127-170.
  • 6§ 493. 1253 Standard: establishment and verification of performance specifications [ EB/OL ]. ( 2003-08- 22 ) [ 2014-06-05 ] http ://www. gpo. gov/fdsys/pkg/ CFR-2003 -title42 -vol3/xml/CFR-2003 -title42-vo13 - part493, xml#seqnum493. 1253.
  • 7Joint Committee for Guides in Metrology. International vocabulary of metrology-Basic and general concepts and associated terms (VIM) [ S ], 3rd edition. JCGM 200 : 2012 ( E/F), JCGM, 2012.
  • 8National Committee for Clinical Laboratory Standards. Evaluation of precision performance of quantitative measurement methods [ S ]. EP05-A2, NCCLS,2004.
  • 9Clinical and Laboratory Standard Institute. User verification of performance for precision and trueness[S]. EP15-A2, CLSI, 2005.
  • 10NetrellaMG.实验统计学[M].毛镇道,蒋子刚,译.上海:上海翻译出版公司,1990:351-356.