摘要
Although the streaked optical pyrometer(SOP)system has been widely adopted in shock temperature measurements,its reliability has always been of concern.Here,two calibrated Planckian radiators with different color temperatures were used to calibrate and verify the SOP system by comparing the two calibration standards using both multi-channel and single-channel methods.A high-color-temperature standard lamp and a multi-channel filter were specifically designed for the measurement system.To verify the reliability of the SOP system,the relative deviation between the measured data and the standard value of less than 5%was calibrated out,which demonstrates the reliability of the SOP system.Furthermore,a method to analyze the uncertainty and sensitivity of the SOP system is proposed.A series of laserinduced shock experiments were conducted at the‘Shenguang-Ⅱ’laser facility to verify the reliability of the SOP system for temperature measurements at tens of thousands of kelvin.The measured temperature of the quartz in our experiments agreed fairly well with previous works,which serves as evidence for the reliability of the SOP system.
Although the streaked optical pyrometer(SOP) system has been widely adopted in shock temperature measurements, its reliability has always been of concern. Here, two calibrated Planckian radiators with different color temperatures were used to calibrate and verify the SOP system by comparing the two calibration standards using both multi-channel and single-channel methods. A high-color-temperature standard lamp and a multi-channel filter were specifically designed for the measurement system. To verify the reliability of the SOP system, the relative deviation between the measured data and the standard value of less than 5% was calibrated out, which demonstrates the reliability of the SOP system.Furthermore, a method to analyze the uncertainty and sensitivity of the SOP system is proposed. A series of laserinduced shock experiments were conducted at the ‘Shenguang-II’ laser facility to verify the reliability of the SOP system for temperature measurements at tens of thousands of kelvin. The measured temperature of the quartz in our experiments agreed fairly well with previous works, which serves as evidence for the reliability of the SOP system.
基金
supported by the National Key R&D Program of China (No. 2017YFA0403200)
the Science Challenge Project (No. TZ2016001)