臂式电子血压计与汞柱式血压计测量结果对比观察探究 被引量:7

Comparative Study on the Measurement Results of Arm-Type Electronic Sphygmometer and Mercury-Type Sphygmometer
下载PDF
导出
摘要 目的对比分析臂式电子血压计测量的血压指标是否高于汞柱式血压计的测量结果并具有临床应用意义。方法选取我院每周一下午专家门诊诊治的高血压患者55例作为研究对象,对研究对象采用泵柱式血压计及臂式电子血压计交替进行血压测量,对照两组血压指标,对比分析数据是否具有显著的统计学差异。结果两组血压检测指标中收缩压的数值存在明显差异(P<0.05),舒张压无明显差异(P>0.05)。结论臂式电子血压计测量的血压数值明显高于泵柱式血压计,其收缩压的数值差异比较明显。 Objective To compare and analyze whether the blood pressure index measured by the arm type electronic sphygmomanometer is higher than that of the mercury sphygmomanometer and has clinical application significance.Methods 55 physical health examiners were selected as the research subjects.The blood pressure measurement was performed alternately with the pump-column sphygmomanometer and the arm-type electronic sphygmomanometer.The blood pressure indicators of the two groups were compared.The comparative analysis data showed significant statistical differences.Results There were significant differences in the values of systolic blood pressure between the two groups(P<0.05),and there was no significant difference in diastolic blood pressure(P>0.05).Conclusion The blood pressure measured by the arm-type electronic sphygmomanometer is significantly higher than that of the pump-column sphygmomanometer.The difference in systolic blood pressure is obvious.
作者 杨成林 郑海艳 熊威伟 YANG Chenglin;ZHENG Haiyan;XIONG Weiwei(Zigui County People’s Hospital,Yichang Hubei 443600,China)
机构地区 秭归县人民医院
出处 《中国医疗设备》 2019年第S01期41-42,共2页 China Medical Devices
关键词 臂式电子血压计 泵柱式血压计 血压数值 arm electronic sphygmomanometer pump column sphygmomanometer blood pressure value
  • 相关文献

参考文献2

二级参考文献52

  • 1向海燕,俞梦孙,刘娟.用有创血压验证示波法测量血压的准确性[J].北京生物医学工程,2005,24(3):195-198. 被引量:9
  • 2张桂平.电子血压计测量原理及存在的问题[J].医疗保健器具,2005(5):48-49. 被引量:16
  • 3贾勇.关于水银式与电子式血压计临床测量血压值偏差的分析[J].中国医疗器械信息,2006,12(3):46-47. 被引量:1
  • 4Aarnes TK,Hubbell JA,Lerche P, et al.Comparison of invasive and oscillometric blood pressure measurement techniques in anesthetized camelids[J].Can Vet J,2012,53(8):881-5.
  • 5Aarnes TK,Hubbell JA,Lerche Eet al.Comparison of invasive and oscillometric blood pressure measurement techniques in anesthetized sheep, goats, and cattle[J].Vet Anaesth Analg,2014,41 (2): 174-85.
  • 6Hatz LA,Hartnack S,K/immerle Let al.A study of measurement of noninvasive blood pressure with the oscillometric device, Sentinel, in isoflurane-anaesthetized horses[J]. Vet Anaesth Analg,2015,42(4):369-76.
  • 7Lee S,Jeon G,Lee G.On using maximum a posteriori probability based on a Bayesian model for oscillometric blood pressure estimation[J].Sensors,2013,13 ( 10): 13609-13623.
  • 8Forouzanfar M,Dajani HR,Groza VZ,et a/.Oscillometric Blood Pressure Estimation: Past, Present, and Future [J].IEEE Rev Biomed Eng,2015,8: 44-6 3.
  • 9Jones DW, Hall JE.Hypertension: Pathways to success[J]. Hypertension,2008,51 (5): 124 - 125 1.
  • 10Soueidan K,Chen S,Dajani HR,et a/.The effect of blood pressure variability on the estimation of the systolic and diastolic pressures[C].IEEE Int,Workshop Medical Measurements and Applications,2010,5:14- 1.

共引文献11

同被引文献45

引证文献7

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部