医用真空清洗机对眼科超声乳化手柄清洗质量效果研究 被引量:6

Study on the Effect of Medical Vacuum Cleaning Machine on the Cleaning Quality of Ophthalmic Ultrasonic Emulsification Handle
下载PDF
导出
摘要 目的探讨医用真空清洗机对眼科超声乳化手柄清洗效果。方法采用YOMA-150医用真空清洗机(减压沸腾)与手工清洗对手术室预处理后的眼科超声乳化手柄进行清洗质量的比较,使用ATP生物荧光法作为判断标准。结果两种清洗方法清洗的超声乳化手柄ATP测试均在正常值,均可适用于手柄的清洗。但两者差异有显著的统计学意义(P<0.001)。结论真空清洗机效果优于手工清洗;清洗时间短,提高了工作效率;手工操作步骤少,降低了职业暴露的风险,减少了器械的损耗;保障了眼科器械的周转和使用。 Objective To explore the effect of medical vacuum cleaner on the cleaning quality of ultrasonic emulsification handles.Methods The cleaning quality of ultrasonic emulsification handles which have been handled in the operating room was compared between YOMA-150 medical vacuum cleaner(decompression boiling)and manual cleaning method.ATP bioluminescence assay was used to judge the effect.Results The ATP test results of the two methods were both at normal values,and they both could be applied to handle the ultrasonic emulsification handles.But there was significant difference between two cleaning methods(P<0.001).Conclusion The cleaning quality of medical vacuum cleaner is better than that of manual cleaning method.It requires a shorter time with improved work efficiency.It has fewer manual steps,which reduces the risk of occupational exposure and the loss of equipment.Medical vacuum cleaner can ensure the turnover and use of the equipment.
作者 景军玲 魏玉芬 夏甜 胥培培 JING Junling;WEI Yufen;XIA Tian;XU Peipei(Shandong Institute of Ophthalmology,Qingdao Shandong 266071,China)
出处 《中国医疗设备》 2019年第S02期1-2,共2页 China Medical Devices
关键词 真空清洗机 手工清洗 超声乳化手柄 清洗质量 medical vacuum cleaner manual cleaning method ultrasonic emulsification handles cleaning quality
  • 相关文献

参考文献3

二级参考文献15

  • 1吴峥峥,瞿佳,樊映川.白内障超声乳化术对角膜内皮细胞的影响[J].国际眼科纵览,2006,30(4):262-265. 被引量:40
  • 2李梅.酶在医疗器械清洗中的效果观察[J].中华医院感染学杂志,2007,17(4):431-432. 被引量:29
  • 3周敏,赵大文.现代医院感染质量管理与控制思路[J].中华医院感染学杂志,2007,17(10):1254-1255. 被引量:42
  • 4Altintas AG, Yilmaz E, Anayol MA, et al. Comparison of corneal edema caused by cataract surgery with different phaco times in dia- betic and non-diabetic patients [ J ]. Ann Ophthalmol, 2006,38 (1) :61-65.
  • 5Raskin E, Paula JS, Cruz AA, et al. Effect of bevel position on the corneal endothelium after phacoemulsification [ J ]. Arq Bras Oftalmol, 2010,73(6) :508-510.
  • 6Faramarzi A, Javadi MA, Karimian F, et al. Corneal endothelial cell loss during phacoemulsification: bevel-up versus bevel-down phaco tip [ J]. J Cataract Refract Surg, 2011,37 ( 11 ) : 1971- 1976.
  • 7Morikubo S, Takamura Y, Kubo E, et al. Corneal changes after small-incision cataract surgery in patients with diabetes mullitus [ ] ]. Arch Ophthalmol, 2004,122(6) :966-969.
  • 8Hugod M, Storr-Paulsen A, Norregaard JC, et al. Corneal endo- thelial cell changes associated with cataract surgery in patients with type 2 diabetes mellitus[J]. Cornea, 2011,30(7) :749-753.
  • 9Wilczynski M, Supady E, Loba P, et al. Comparison of early cor- neal endothelial cell loss after coaxial phacoemulsification through 1.8 mm microincision and bimanual phacoemulsification through 1.7 mm microincision [ J ]. J Cataract Refract Surg, 2009,35 (9) : 1570-1574.
  • 10Zafar A,Habib F,Hadwani R,et al. Impact of infection con- trol activities on the rate of needle stick injuries at a tertiary care hospital of pakistan over a period o six years: an obser rational study[J]. BMC Infectious Diseases, 2009,9 : 78.

共引文献17

同被引文献80

引证文献6

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部