摘要
针对JJF 1366-2012《温度数据采集仪校准规范》在日常校准过程中所遇到的四个问题:温湿度数据采集仪温度部分的校准;校准与使用状态不一致对测量结果的影响;校准温度数据采集仪用特制容器设计及温度验证;温度采集仪记录数据的时间间隔影响等进行探讨,并对以上问题提出一些解决方法并予以验证。
In view of the four problems encountered in the daily calibration process of the Temperature Data Collector Calibration Specification JJF1366-2012:calibration of the temperature part of the temperature&humi data collector,the effect of inconsistent calibration and use status on the measurement results,the use of stainless steel tank and temperature verification for auxiliary calibration,and the interval between the recording of data by the temperature collector.Propose feasible methods and verify them.
作者
胡健
徐标
Hu Jian;Xu Biao(South China National Centre of Metrology,Guangdong Guangzhou 510405)
出处
《电子质量》
2020年第1期59-61,共3页
Electronics Quality
关键词
温度数据采集仪
时间间隔
温湿度数据采集仪
特制容器设计
Temperature Data Acquisition Instruments
Interval
Temperature and Humidity Data Acquisition Instruments
Special container design