Pentacam眼前节分析仪和Keratron Scout角膜地形图仪测量近视患者角膜前表面曲率的重复性和一致性分析 被引量:2

Comparison of keratometry consistency between Pentacam and Keratron Scout topography among myopic patients
下载PDF
导出
摘要 目的探讨以Schiempflug原理设计的Pentacam眼前节分析仪和以Placido盘为基础的Keratron Scout角膜地形图仪测量角膜前表面曲率的重复性,并分析两种仪器测量结果的一致性。方法随机数字表法随机选取于2018年1月1~30日在我院眼视光中心行角膜激光近视矫正术前检查的69例患者,所有受试者由同一检查者分别采用Pentacam眼前节分析仪和Keratron Scout角膜地形图仪重复测量3次,两种仪器测量顺序随机,记录患者平坦轴角膜曲率(Kf)、陡峭轴角膜曲率(Ks)和平均角膜曲率(Km),采用变异系数(CV%)、组内相关系数(ICC)(95%CI)和Cronbach’s Alpha系数评价两种仪器测量结果的可重复性,t检验比较两种仪器测量结果的差异,Pearson相关分析法分析其相关性;Bland-Altman图分析一致性界限范围。结果3次重复测量时,Pentacam眼前节分析仪测量Kf的CV为0.39%,ICC(95CI%)为0.811(0.651~0.907),Cronbach’s Alpha系数为0.935,Ks的CV为0.28%,ICC为0.975(0.959~0.985),Cronbach’s Alpha系数为0.991,Km的CV为0.25%,ICC为0.944(0.911~0.967),Cronbach’s Alpha系数为0.981;Keratron Scout角膜地形图仪测量Kf的CV为0.43%,ICC为0.820(0.678~0.951),Cronbach’s Alpha系数为0.967,Ks的CV为0.47%,ICC为0.937(0.908~0.959),Cronbach’s Alpha系数为0.978,Km的CV为0.39%,ICC为0.937(0.907~0.958),Cronbach’s Alpha系数为0.978。Pentacam眼前节分析仪和Keratron Scout角膜地形图仪测量Kf、Ks和Km的结果差异无统计学意义(Kf:P=0.187;Ks:P=0.721;Km:P=0.169),两种仪器测得结果有相关性(Kf:r=0.622,P=0.000;Ks:r=0.980,P=0.000;Km:r=0.983,P=0.000),Kf的95%Lo Al为-0.43~0.40D,Ks的95%Lo Al为-0.56~0.59 D,Km的95%LoAl-0.46~0.39 D。结论以Schiempflug原理设计的Pentacam眼前节分析仪和以Placido盘为基础的Keratron Scout角膜地形图仪测量角膜屈光手术患者角膜前表面曲率均具有良好的可重复性,两种仪器测量得角膜前表面曲率结果准确,一致性好,可以相互验证。 Objective To evaluate the consistency of keratometry measurement by Pentacam Front Section Analyzer(schiempflug image)and Keratron Scout Corneal Topography(Placido disc)and analyze the repeatability of each instrument.Methods Sixty nine patients who underwent corneal refractive surgery from January 1 st to 30 th at the ophthalmology department of Leshan people’s hospital were randomly selected using random number table method preoperatively.All patients were examined by the same doctor using pentacam and keratron scout corneal topography respectively.Three consecutive measurements were taken using each instrument for repeatability analysis,and the sequence of measurement by these two instruments were also random.The flat keratometry(Kf),steep keratometry(Ks)and mean keratometry(Km)were recorded and compared using coefficient of variation,intraclass correlation coefficient(ICC),and Cronbach’s Alpha Coefficient.T test was used to assess the difference in the measurements by these two instruments,Pearson correlation analysis was used for correlation analysis and Bland-Aaltman analysis was used for consistency analysis.Results Both instruments showed good intra-group repeatability with low coefficient of variation,high ICC and Cronbach’s Alpha coefficient.Pearson correlation analysis showed significant correlations between the results(Kf r=0.622,P=0.000;Ks:r=0.980,P=0.000;Km:r=0.983,P=0.000).No significant difference was found between Kf,Ks and Km measured by these two instruments.95%Lo Al of the two instruments was-0.43 to 0.40 D,-0.56 to 0.59 D and-0.46 to 0.39 D for Kf,Ks and Km by Bland-altman analysis method,respectively.Conclusions Pentacam Front Section Analyzer(Schiempflug)and Keratron Scout corneal topography(Placido disc)showed good repeatability and consistency in determining Kf,Ks and Km of the anterior corneal curvature in myopic patients preoperatively.The two instruments could be used for mutual corroboration.
作者 郝更生 赵姣 曾莉 刘宗顺 胡卜丹 毛科 Hao Gengsheng;Zhao Jiao;Zeng Li;Liu Zongshun;Hu Budan;Mao Ke(Department of Ophthalmology,Leshan People’s Hospital,Leshan Ophthalmic Center,Leshan 614000,China)
出处 《临床眼科杂志》 2020年第1期58-61,共4页 Journal of Clinical Ophthalmology
关键词 Pentacam眼前节分析仪 Keratron Scout角膜地形图仪 角膜曲率 一致性 重复性 Pentacam Front Section Analyzer Keratron Scout Corneal Topography Keratometry Consistency Repeatability
  • 相关文献

参考文献3

二级参考文献53

  • 1Bogan SJ, Waring GO 3d, Ibrahim O, et al. Classification of normal corneal topography based on computer-assisted videokeratography. Arch Ophthalmol 1990 ; 108:945-949.
  • 2Mejia-Barbosa Y, Malacara-Hernandez D. A review of methods for measuring corneal topography. Optom Vis Sei 2001 ;78 (4) : 240-253.
  • 3Lackner B, Schmidinger G, Skorpik C. Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci 2005 ;82(9) :858-861.
  • 4Holladay JT, Belin MW, Chayet AS, et al . Next-generation technology for the cataract and refractive surgery.Cat Ref Surg Today Suppl 2005;1:1-11.
  • 5Kent C. The anterior chamber from every angle.Rev Opbthalmol 2005 ; 12(6) :33-38.
  • 6Gardner BG, Taub MB. New diagnostic devices in Eye care.Rev Optom 2004 ; 15:79-85.
  • 7Roberts C. Characterization of the inherent error in a spherically- biased corneal topography system in mapping a radially aspheric surface. J Refract Corneal Surg 1994;10(2) :103-111.
  • 8Belin MW, Khachikian SS. Highlights of Ophthalmology, City of Knowledge. Am J Opbthalmol 2007 ; 143 ( 3 ) :500-503.
  • 9Bogan SJ, Waring GO 3d, Ibrahim O, et al . Classification of normal corneal topography based on computer-assisted videokeratography. Arch Ophthalmol 1990 ; 108 ( 7 ) :945-949.
  • 10Naufal SC, Hess JS, Friedlander MH, et al. Rasterstereography-based classification of normal eomeas.J Cataract Refract Surg 1997 ;23(2) :222-230.

共引文献25

同被引文献18

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部