二维矩阵电离室验证Acuros BV算法计算相对剂量分布的可行性研究 被引量:2

Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV
原文传递
导出
摘要 目的研究PTW729二维矩阵电离室验证Acuros BV算法在相对剂量分布计算方法的可行性,分析该方法的优势和劣势,为实际临床工作提供参考。方法基于自建测量模体,在相同计划和测量条件下,使用PTW729和胶片方法,分别测试相同位置处的剂量分布,对两种方法的测量结果和Acuros BV算法计算结果分别进行γ比对分析,并测试PTW729方法的稳定性。结果PTW729和胶片的测试结果与Acuros BV算法计算结果的γ比对值分别为98.9%、95.9%,PTW729与胶片测试结果的γ比对值为88.0%。PTW729稳定性测试实验γ比对值分别为95.0%、100.0%、100.0%。结论PTW729二维矩阵电离室可应用于Acuros BV算法相对剂量分布的快速验证。 Objective To study the feasibility of using the PTW7292D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm.Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices.Methods Based on self-built measurement phantoms,the dose distribution on the same slice of the phantom was measured with PTW729 and film,respectively,under the same measurement condition and plan.The dose distributions obtained by the two method were compared with the result calculated with Acuros BV,separately,by usingγanalytical tool.And the stability of the PTW729 was tested.Results Theγcomparison value was 95.9%between the film and Acuros BV,98.9%between the PTW729 and Acuros BV and 88.0%between the film and PTW729,with 95.0%,100.0%,and 100.0%,in their stability test respectively.Conclusions PTW7292D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution.
作者 赵强 吴湘阳 常晓斌 冯涛 杨迪 张坤 屈喜梅 王学敏 邓佳 Zhao Qiang;Wu Xiangyang;Chang Xiaobin;Feng Tao;Yang Di;Zhang Kun;Qu Ximei;Wang Xuemin;Deng Jia(Department of Radiotherapy,Shaanxi Provincial Cancer Hospital,Xi′an 710061,China)
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2020年第1期59-63,共5页 Chinese Journal of Radiological Medicine and Protection
关键词 二维矩阵电离室 Acuros BV 近距离治疗 2D array ion-chamber Acuros BV Brachy therapy
  • 相关文献

参考文献3

二级参考文献41

  • 1李文华,张书旭,徐海荣.宫颈癌后装治疗的剂量分布测试[J].中国医学物理学杂志,2005,22(3):501-504. 被引量:3
  • 2刘孜,罗伟,王国庆,王蕊华,郑伟,杨晓娟,高彩霞,王娟.中晚期宫颈癌后装腔内放疗剂量分布影响因素探讨[J].中华放射肿瘤学杂志,2006,15(6):444-447. 被引量:15
  • 3李浦,狄小云,余华,杨丽芳.宫颈癌后装治疗中直肠、膀胱剂量的研究[J].中国医学物理学杂志,2007,24(4):243-244. 被引量:4
  • 4Haie-Meder C, Potter R, Van Limbergen E, et al. Recom?mendations from Gynaecological (GYN) GEC-ESTRO Work?ing Group (I): concepts and terms in 3D image-based 3D treatment planning in cervix cancer brachytherapy with em?phasis on MRI assessment of GTV and CTV[J]. Radiother Oncol, 2005, 74: 235-245.
  • 5NathR, AndersonLL, LuxtonG, etal' Dosimetryofintersti?tial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group[J]. Med Phys, 1995,22: 209-234.
  • 6Niroomand-Rad A, Blackwell CR, Coursey BM, et al. Ra?diochromic film dosimetry: Recommendations of AAPM Radi?ation Therapy Committee Task Group 55[J]. Med Phys, 1998, .25: 2093-2115.
  • 7HerzenJ, Todorovic M, Cremers F, et al. Dosimetric evalua?tion of a 2D pixelionization chamber for implementation in clin?ical routine[J]. Phys Med BioI, 2007,52: 1197-1208.
  • 8Saminathan S, Manickan R, Chandraraj V, et al. Dosimetric study of 2D ion chamber array matrix for the modem radio?therapy treatment verification[J].J Appl Clin Med Phys , 2010, 11: 116-127.
  • 9Wolfsberger LD, Wagar M, Nitsch P, et al. Angular dose de?pendency of MatriXXTM and its calibration[J].J Appl Clin Med Phys, 2010, 11: 241-251.
  • 10Dobler B, Streck N, Klein E, et al. Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX-a feasibility study[J]. Phys Med BioI, 2010, 55: N39-N55.

共引文献16

同被引文献11

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部