期刊文献+

固体进样汞镉测定仪同时测定化妆品中的汞和镉 被引量:3

Simultaneous Determination of Mercury and Cadmium in Cosmetics by Solid Sampling Atomic Fluorescence Spectrometry
下载PDF
导出
摘要 本文建立了采用固体进样原子荧光汞镉测定仪同时测定化妆品中汞和镉的快速分析方法,确定了空气流量0.4 NL/min,氩氢混合气流量1.0 NL/min,镉蒸发功率80 W,镉释放功率30 W,汞灰化功率240 W,汞释放功率170 W的条件,并以此条件对样品进行测定。该方法的检出限汞为0.0018 ng、镉为0.0015 ng,回收率为94.7%~107.7%,相对标准偏差(RSD)小于6%,分析时间小于5 min。 This paper establishes a rapid method for the determination of mercury and cadmium in cosmetics by using a solid sampling atomic fluorescence spectrometry.The conditions of air flow rate is 0.4 NL/min,argon hydrogen mixture flow rate is 1.0 NL/min,cadmium evaporation power is 80 W,cadmium release power is 30 W,mercury ashing power is 240 W,mercury release power is 170 W,and the samples are analyzed under these conditions.The detection limit of the method is 0.0018 ng for mercury and 0.0015 ng for cadmium,the recovery is 94.7%-107.7%,RSD is less than 6%,and the analysis time is less than 5 min.
作者 林隆强 张敏 张燕燕 陈德平 LIN Longqiang;ZHANG Min;ZHANG Yanyan;CHEN Deping(Shantou Customs District,Shantou,Guangdong,515041,China)
机构地区 汕头海关
出处 《检验检疫学刊》 2020年第1期18-21,共4页 Journal of Inspection and Quarantine
基金 广东检验检疫局科技计划项目:新型固体进样汞镉测定仪检测化妆品中的汞、镉及其健康风险评估(2018GDK53)。
关键词 固体进样 汞镉同测 化妆品 Solid Sampling Simultaneous Determination of Mercury and Cadmium Cosmetics
  • 相关文献

参考文献5

二级参考文献31

共引文献41

同被引文献22

引证文献3

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部