期刊文献+

SKALAR连续流动分析仪测定地表水中的挥发酚 被引量:5

SKALAR continuous flow analyzer for determination of volatile phenols in surface water
下载PDF
导出
摘要 本文通过连续流动分析仪(荷兰SKALAR San++)对地表水中的挥发酚的测定,对校准曲线、加标回收率、相对误差、准确度和精确度等方面进行探究。实验结果表明在浓度范围为20~100 ug/L,挥发酚线性良好,相关系数为r=0.9998,样品加标回收率范围为93.0%~102.0%,当浓度为20.0,60.0 ug/L时,相对误差范围分别为-1.0%~5.5%,-2.3%~3.8%。得到连续流动分析法测定地表水中挥发酚结果是令人满意的,能满足大批量的环境样品快速分析。 In this paper,a continuous flow analyzer (SKALAR San ++ in the Netherlands) was used to determine volatile phenols in surface water,and the calibration curve,standard recovery,relative error,accuracy,and precision were investigated.The experimental results show that in the concentration range of 20~100 ug / L,the volatile phenols have good linearity,the correlation coefficient is r=0.9998,and the sample spiked recovery range is 93.0%~102.0%.When the concentration is 20.0,60.0 ug / L,the relative The error ranges are-1.0% ~ 5.5% and-2.3% ~ 3.8%.The continuous flow analysis method is satisfactory for the determination of volatile phenols in surface water,and can meet the rapid analysis of large quantities of environmental samples.
作者 夏文盛 Xia Wensheng(Nanping Jianyang Environmental Monitoring Station,Nanping Fujian 354200,China)
出处 《环境与发展》 2020年第1期172-173,共2页 Environment & Development
关键词 挥发酚 连续流动分析 快速测定 Volatile phenol Continuous flow analysis Rapid determination
  • 相关文献

参考文献1

二级参考文献1

共引文献10

同被引文献36

引证文献5

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部