摘要
文章介绍了傅立叶变换红外光谱仪测定润滑油中抗氧剂2,6-二叔丁基对甲酚(T501)的含量,考察不同光程长的液体池对抗氧剂T501最低检测限的影响,同时确定抗氧剂T501的最低检测限:光程长为3 mm时,抗氧剂T501的最低检测限最低,达到0. 01%(质量分数),相关系数为0. 9984。
The detection limit of 2,6-di-tert-butyl-4-methylphenol determined by Fourier Transform Infrared Spectrometer was introduced in this paper. The effect of liquid pool with different optical path length on the minimum detection limit of antioxidant(T501)was investigated,the minimum detection limit of antioxidant(T501) was determined. When the optical path length is 3 mm,the lowest detection limit of antioxidant(T501) is 0. 01%(mass fraction),and the correlation coefficient is 0. 9984.
作者
乔琦
刘燕
王储
张向英
QIAO Qi;LIU Yan;WANG Chu;ZHANG Xiang-ying(CNOOC Oil&Gas(Taizhou)Petrochemical CO.,LTD.,Taizhou 225300,China)
出处
《润滑油》
CAS
2020年第1期37-39,共3页
Lubricating Oil
关键词
傅立叶变换红外光谱仪
抗氧剂
2
6-二叔丁基对甲酚
检测限
Fourier Transform Infrared Spectrometer
antioxidant
2,6-di-tert-butyl-4-methylphenol
detection limit