期刊文献+

西门子MAGNETOM Avanto 1.5T磁共振维修三例 被引量:6

Three Cases of Siemens MAGNETOM Avanto 1.5T Magnetic Resonance Imaging Equipment Maintenance
下载PDF
导出
摘要 长期高负荷的使用导致磁共振设备故障发生率逐渐升高,对临床工程师迅速排除故障保障设备正常运行提出了更高的要求。本文对我院西门子MAGNETOM Avanto 1.5T磁共振设备发生过的发射器和RFSU电源故障、光纤断裂故障和PF信号线故障三例故障现象,逐步分析故障原因,通过更换Transmitter板和RFSU电源、光纤、PF信号线使设备恢复正常运行。本文旨在通过对这三个案例的分享,加强与同行的交流和学习,共同提高设备保障能力。 Long-term and high-load use has gradually increased the incidence of magnetic resonance equipment failures,which has put forward higher requirements for clinical engineers to quickly troubleshoot and ensure the normal operation of the equipment.In this paper,three failures of the Siemens MAGNETOM Avanto 1.5T magnetic resonance imaging equipment,including transmitter and rfsu power supply failure,optical fiber break failure and PF signal line failure,were analyzed step by step.Then,the normal operation of the equipment was restored by replacing the Transmitter board and RFSU power supply,optical fiber and PF signal line.By sharing these three cases,we hope to strengthen communication and learning with peers and jointly improve equipment support capabilities.
作者 王子洪 郭海涛 陈星 周翔 韩锦川 WANG Zihong;GUO Haitao;CHEN Xing;ZHOU Xiang;HAN Jinchuan(Department of Medical Engineering,The First Hospital Affiliated to Army Medical University,Chongqing 400038,China;Department of Radiology,The First Hospital Affiliated to Army Medical University,Chongqing 400038,China)
出处 《中国医疗设备》 2020年第4期173-175,共3页 China Medical Devices
关键词 1.5T磁共振 射频放大器 光纤断裂 PF信号线故障 1.5T magnetic resonance imaging radio frequency amplifier fiber break PF signal line failure
  • 相关文献

参考文献14

二级参考文献49

共引文献65

同被引文献51

引证文献6

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部