Sysmex XN-3000血液分析仪检测外周血有核红细胞的临床价值评估 被引量:3

Evaluation of the Clinical Value of Sysmex Xn-3000 Blood Analyzer in Detecting Peripheral Blood Nucleated Red Blood Cells
下载PDF
导出
摘要 目的:评估Sysmex XN-3000血液分析仪检测外周血有核红细胞的临床价值。方法:选取2017年1月~2018年7月在本院住院患者80例,对患者血液标本均进行Sysmex XN-3000血液分析仪检测及显微镜镜检法,观察检测结果。结果:将显微镜镜检法作为参照标准,Sysmex XN-3000检测外周血有核红细胞阴性28例,血片手工复检1例外周血有核红细胞阳性,Sysmex XN-3000检测外周血有核红细胞阳性52例,血片手工复检50例阳性,Sysmex XN-3000检测敏感性98.04%(50/51),特异性93.10%(27/29)。结论:Sysmex XN-3000血液分析仪检测外周血有核红细胞,操作简便快捷,具有较高的准确性。 Objective:To evaluate the clinical value of Sysmex XN-3000 blood analyzer in detecting peripheral blood nucleated red blood cells.Methods:80 inpatients from our hospital(from January 2017 to July 2018)were selected,and blood samples of the patients were tested by Sysmex XN-3000 blood analyzer and microscopic examination to observe the test results.Results:Using microscopy as a reference standard,Sysmex XN-3000 detected peripheral blood nucleated red blood cells in 28 cases,and blood samples were manually re-examined with one exception,and peripheral blood nucleated red blood cells were positive.Fifty-two patients were positive by manual blood test.The sensitivity of Sysmex XN-3000 was 98.04%(50/51)and the specificity was 93.10%(27/29).Conclusion:The Sysmex XN-3000 blood analyzer can detect nucleated red blood cells in peripheral blood.The operation is simple and fast,and it has high accuracy.
作者 许娇 XU Jiao(Department of Laboratory Medicine,Panjin Central Hospital,Liaoning Panjin 124010)
出处 《中国医疗器械信息》 2020年第7期140-141,共2页 China Medical Device Information
关键词 SYSMEX XN-3000血液分析仪 外周血 有核红细胞 实验室诊断 sysmex XN-3000 blood analyzer peripheral blood nucleated red blood cells laboratory diagnosis
  • 相关文献

参考文献5

二级参考文献48

  • 1中国实验室国家认可委员会技术委员会医学分委会.医学实验室质量管理与认可指南[M].北京:中国计量出版社,2004.
  • 2王欣,滕娟,郝少丽.血细胞分析仪计数幼稚红细胞的准确性评价[J].中华检验医学杂志,2007,30(9):1056-1057. 被引量:7
  • 3中华人民共和国卫生部.WS/T347-2011:血细胞分析的校准指南[S].北京:中国标准出版社,2011.
  • 4Harrison P, Horton A, Grant D, et al. Immunoplatelet count- ing: a proposed new reference procedure[J]. Brit J Heamat, 2000,108:228-235.
  • 5England J M, Rowan R M, Bins M, et al. Recommended methods for the visual determinat ion of white cell and plat eletcount[Z]. 1988/WHO-LAB-88.3.
  • 6International Council for Standardization in Haematology Ex- pert Panel on Cytometry, International Society of Laboratory Hematology Task Force on Platelet Counting. Platelet count- ing by the RBC/platelet ratio method. A reference method[J]. Am J Clin Pathol, 2001,115(3):460-464.
  • 7Shitara M, Umezu S, Katsuno H. Effects of the ISO15189 ac- creditation on clinical laboratory[J]. Rinsho Byori, 2009,57(6): 521-526.
  • 8Okubo S, Yokota H, lkeda H, et al. Overview of ISO15189 and accreditation with it[J]. Rinsho Byori, 2010,58(7):676-684.
  • 9Briggs C, Harrison P, Grant D, et al. New quantitative param- eters on a recently introduced automated blood cell counter- the XE-2100[J]. Clin Lab Haematol, 2000,22(6):345-350.
  • 10叶应妩,王毓三.全国临床检验操作[M].南京:东南大学出版社,1991:13.

共引文献20

同被引文献28

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部