Philips CT AURA是单排CT中设计经典的设备。本文介绍了CT的工作原理,并结合实际维修工作进行分析Philips AURA螺旋CT的通讯故障与高压故障。通过查看报警信息,对照维修手册,对发生的故障进行查找故障原因,并通过加固CPU板、更换阴阳极模块使CT设备恢复正常使用。另外,临床工程师还应该加强设备巡检以及增加预防性维护的频率,延长设备的使用寿命,为医院节约开支。
Philips CT AURA is a classic equipment designed in single row CT.The working principle of CT was introduced,and the communication fault and high voltage fault of Philips AURA spiral CT were analyzed based on the actual maintenance work in this paper.The cause of the fault was found by checking the alarm information and comparing with the maintenance manual,and the CT equipment was restored to normal use by strengthening the CPU board and replacing the anode and cathode modules.In addition,the clinical engineer should also strengthen the equipment inspection and increase the frequency of preventive maintenance,and extend the service life of the equipment,so as to save costs for the hospital.
作者
刘春涛
娄睿
赵建平
LIU Chuntao;LOU Rui;ZHAO Jianping(Department of Ultrasonic,Liaocheng Dongchangfu District Maternal and Child Health Care Center,Liaocheng Shandong 252000,China)
出处
《中国医疗设备》
2020年第5期181-183,186,共4页
China Medical Devices
关键词
螺旋CT
通讯故障
高压故障
图像倾斜
spiral CT
communication failure
high voltage fault
image tilt