期刊文献+

4008B型血液透析机电源故障维修案例分析 被引量:2

A case analysis:Maintenance of power supply failure of Fresenius 4008B hemodialysis machine
下载PDF
导出
摘要 目的:分析血液透析机电源故障案例,提高医疗设备元器件级的维修能力。方法:对4008B型血液透析机电源结构和电源模块故障进行分析,逐步排查故障点。结果:经对电源电路板分析排查,确定故障为开关电源管理芯片TOP246YN芯片损坏,更换同型号元器件后故障排除。结论:在医疗设备维修中,积极尝试拆机检修,逐级分析判断,结合资料查询,可开展元器件级的维修。 Objective:To analyze the case of power supply failure of hemodialysis machine so as to improve the maintenance ability for the component level of medical equipment.Methods:The structure and the module fault of the power supply of Fresenius 4008 B hemodialysis machine were analyzed and the failure point was checked step by step.Results:Through analysis and troubleshooting of the power supply circuit board,the failure was confirmed to be the damage of one TOP246 YN chip,which is a management chip of switching power.The fault was removed after the failure component was replaced by the same type of component.Conclusion:In medical equipment maintenance,hospital should positively try to implement disassemble maintenance,gradual analysis and judgment and inquiry with data.And the component-level maintenance should be carried out.
作者 高婧颖 陆银春 戴京 邓勇 GAO Jing-ying;LU Yin-chun;DAI Jing(Medical Equipment Department,Wuxi No.2 People’s Hospital,Wuxi 214002,China;不详)
出处 《中国医学装备》 2020年第5期219-221,共3页 China Medical Equipment
基金 湖南省教育厅课题资助(19c1722)“银杏黄酮、银杏内酯在治疗心脑血管疾病中用药合理性研究” 郴州市静脉药物调配技术研发中心项目资助(yfzx201911)“郴州市静脉药物调配技术研发中心”。