使用不同眼压计比较翼状胬肉术后绷带镜对眼压的影响 被引量:1

Comparative measurement of intraocular pressure by Rebound and non-contact tonometry in patients wearing bandage contact lenses after pterygium excision
下载PDF
导出
摘要 目的:利用回弹式眼压计(RT)及非接触式眼压计(NCT)测量并比较翼状胬肉术后配戴治疗用绷带镜时与裸眼眼压的差异。方法:选取2019-04/06在开封市中心医院开封眼病医院手术治疗的翼状胬肉患者(鼻侧原发性翼状胬肉小于4mm且术后配戴治疗用绷带镜)50例50眼。于术后14d用RT及NCT分别在戴镜状态、裸眼状态测量患者眼压,比较并分析翼状胬肉术后治疗用绷带镜对眼压测量的影响。结果:RT在戴镜状态、裸眼状态测得的眼压分别为18.20±3.19、15.17±3.80mmHg,相差3.04±1.79mmHg,且前后两次测量具有良好的相关性(r=0.884,P<0.001);NCT在戴镜状态、裸眼状态测得的眼压分别为15.74±3.23、13.19±3.89mmHg,相差2.55±1.88mmHg,且具有良好的相关性(r=0.876,P<0.001);在戴镜状态及裸眼状态,两种设备测量所得眼压分别相差2.46±1.45、1.98±1.67mmHg,RT测量结果均显著高于NCT测量结果(P<0.001)。结论:翼状胬肉术后应用RT、NCT测量配戴治疗用绷带镜状态与裸眼状态的眼压一致性均较好,但在配戴治疗用绷带镜时两种眼压计均有高估患者眼压风险,且应用RT在戴镜状态及裸眼状态眼压高估较NCT更明显。 ·AIM:To compare the measurement of intraocular pressure(IOP)through a bandage contact lens with the“native”measurement by rebound tonometry and non-contact tonometry in patients after pterygium excision.·METHODS:Fifty consecutive patients(50 eyes)undergone pterygium excision(unilateral nasal,primary pterygium,horizontal length<4mm),and conjunctival autografting were included in this prospective study.IOP measurements were obtained by Rebound tonometry and non-contact tonometry in random order with(lens measurement)and without(native measurement)a bandage contact lens half a month after operation.We compared the mean values(validity parameter)and standard deviation(precision parameter)of the two individual measurements in each case using the paired t-test 14d after surgery.·RESULTS:With the rebound tonometry we detected statistically significant higher values in the contact lens measurements(18.20±3.19 vs 15.17±3.80mmHg in the native measurements;P<0.001),a good correlation with r=0.884 and mean difference was 3.04±1.79mmHg;With the non-contact tonometry we detected statistically significant higher values in the contact lens measurements(15.74±3.23 vs 13.19±3.89mmHg in the native measurements;P<0.001),a good correlation with r=0.876 and mean difference was 2.55±1.88mmHg.In the contact lens measurements and native measurements,we detected statistically significant higher values by Rebound tonometry than that by non-contact tonometry(P<0.001),and mean difference was 2.46±1.45mmHg,1.98±1.67mmHg.·CONCLUSION:The use of rebound tonometry and non-contact tonometry shows good consistency between lens measurement and native measurement.However,it should be noted that the average of the measurements over contact lens by rebound tonometry and non-contact tonometry were found to be higher than that in native measurement,and the average of the measurements with and without lens by rebound tonometer was found to be higher than what was measured by non-contact tonometry.
作者 黄明可 任保童 任艳竹 孟凡毅 皮百木 Ming-Ke Huang;Bao-Tong Ren;Yan-Zhu Ren;Fan-Yi Meng;Bai-Mu Pi(Department of Ophthalmology, Kaifeng Central Hospital, Kaifeng 475000, Henan Province, China)
出处 《国际眼科杂志》 CAS 北大核心 2020年第6期1091-1094,共4页 International Eye Science
关键词 翼状胬肉切除手术 治疗用绷带镜 眼压 回弹式眼压计 非接触眼压计 pterygium excision bandage contact lens intraocular pressure rebound tonometry non-contact tonometry
  • 相关文献

参考文献3

二级参考文献34

  • 1姚广鹏,晋秀明.角膜绷带镜治疗糖尿病性视网膜病变患者玻切术后持续性角膜上皮缺失治疗的临床观察[J].中国急救医学,2017,37(A01):245-246. 被引量:3
  • 2Ping-Bo Ouyang,Cong-Yi Li,Xiao-Hua Zhu,and Xuan-Chu Duan.Assessment of intraocular pressure measured by Reichert Ocular Response Analyzer,Goldmann Applanation Tonometry,and Dynamic Contour Tonometry in healthy individuals[J].International Journal of Ophthalmology(English edition),2012,5(1):102-107. 被引量:4
  • 3王艳,徐建江,孙兴怀.大泡性角膜病变病因分析[J].中国眼耳鼻喉科杂志,2005,5(6):363-364. 被引量:13
  • 4Kyoung Nam Kim,Jin Wook Jeoung,Ki Ho Park,Min Kyu Yang,Dong Myung Kim.Comparison of the new rebound tonometer with Goldmann applanation tonometer in a clinical setting[J]. Acta Ophthalmologica . 2013 (5)
  • 5Meghan S. Flemmons,Ya-Chuan Hsiao,Jacqueline Dzau,Sanjay Asrani,Sarah Jones,Sharon F. Freedman.Home Tonometry for Management of Pediatric Glaucoma[J]. American Journal of Ophthalmology . 2011 (3)
  • 6Jorge Manuel Martins Jorge,Jose M. González-Méijome,Antonio Queirós,Paulo Fernandes,Manuel A. Parafita.Correlations Between Corneal Biomechanical Properties Measured With the Ocular Response Analyzer and ICare Rebound Tonometry[J]. Journal of Glaucoma . 2008 (6)
  • 7Javier Moreno-Monta?és,Noelia García,Ana Fernández-Hortelano,Alfredo García-Layana.Rebound Tonometer Compared With Goldmann Tonometer in Normal and Pathologic Corneas[J]. Cornea . 2007 (4)
  • 8Alexandra Anton,Matthias Neuburger,Daniel B?hringer,Jens F. Jordan.Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses[J]. Graefe’s Archive for Clinical and Experimental Ophthalmology . 2013 (7)
  • 9Ioannis Halkiadakis,Aimilianos Stratos,George Stergiopoulos,Eleni Patsea,Sotiris Skouriotis,Panagiotis Mitropoulos,Dimitrios Papaconstantinou,Gerasimos Georgopoulos.Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects[J]. Graefe’s Archive for Clinical and Experimental Ophthalmology . 2012 (8)
  • 10Jose M.Martinez‐de‐la‐Casa,MariaJimenez‐Santos,FedericoSaenz‐Frances,MariaMatilla‐Rodero,CarmenMendez‐Hernandez,RocioHerrero‐Vanrell,JulianGarcia‐Feijoo.Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice[J]. Acta Ophthalmologica . 2011 (7)

共引文献25

同被引文献7

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部