期刊文献+

基于起始平衡法气体活塞压力计有效面积的测量及其影响因素探讨 被引量:3

The Effective Area Measurement of Gas Piston Manometer Based on Initial Equilibrium Method and Its Iinfluencing Factors Are Discussed
下载PDF
导出
摘要 依据JJG 1086-2013《气体活塞式压力计》检定规程,基于起始平衡法对0.01级气体活塞压力计的活塞有效面积测量过程进行了论述,研究各个因素对测量结果的影响,依此进行判断和分析,推导出影响气体活塞压力计活塞有效面积测量结果扩展不确定度的各个分量公式,并以此计算出分量大小,为减小测量结果的扩展不确定度提出建议。 According to JJG 1086-2013"Pneumatic Piston Gauge",based on the initial balance method,this paper discusses the measurement process of the effective area of the piston of 0.01 pneumatic piston gauge,studies the influence of various factors on the measurement results,makes judgment and analysis based on this,and deduces the various component formulas affecting the expanded uncertainty of the measurement results of the effective area of the piston of the pneumatic piston gauge based on this,the size of the component is calculated,and suggestions are put forward to reduce the expanded uncertainty of the measurement results.
作者 孙涛 俞和君 林雁波 陈庆峰 SUN Tao;YU Hejun;LIN Yanbo;CHEN Qingfeng
出处 《计量与测试技术》 2020年第6期71-74,77,共5页 Metrology & Measurement Technique
关键词 气体活塞式压力计 有效面积 不确定度 pneumatic piston gauge effective area expanded uncertainty
  • 相关文献

参考文献5

二级参考文献13

  • 1李燕华.活塞式压力计专用砝码质量在计算和修正中的问题[J].计量技术,2007(4):50-52. 被引量:15
  • 2JJG 59-2007活塞式压力计[S].北京:中国计量出版社,2007.
  • 3R S Dadson.The Pressure Balance Theory and Practice[M].London:Her Majesty's Stationery Office,1982.
  • 4国际法制计量组织.OIMLR110-1994,压力天平国际建议[S].1994.
  • 5国家质量监督检验检疫总局.JJG59-2007,活塞式压力计检定规程[S].2007.
  • 6国家计量总局.质量计量基础[M].北京:技术标准出版社,1982.
  • 7许第昌.压力计量测试[M].北京:中国计量出版社.1986.
  • 8JJG59—2007《活塞式压力计》检定规程.北京:中国计量出版社,2007.
  • 9JJFl059—1999《测量不确定度评定与表示》.北京:中国计量出版社,1999.
  • 10《计量技术基础》.北京:原子能出版社,2002.