相同存储条件下三种全血糖化血红蛋白A1c检测系统检测结果对比

Comparison of Detection Results of Three Kinds of Total Glycated Hemoglobin A1c Detection Systems Under the Same Storage Conditions
下载PDF
导出
摘要 目的:比较相同存储条件下三种全血糖化血红蛋白A1c检测系统检测结果,进一步探究不同检测方法的稳定性。方法:选取2018年7月~2019年7月在本院接受的体检人群50例,选取体检人群清晨空腹以及饭后2h的2份血液样本,注入乙二胺四乙酸二钾(EDTA-K2)抗凝试管中,随后用Bio-Rad variant Ⅱ Turbo、Primus UItra2以及Roch Modular PPI三种检测方法检测,比较三种检测方法结果的差异性以及稳定性。结果:三种检测方法检测结果比较不具有差异性(P>0.05);在检测结果稳定性方面,Bio-Rad variant Ⅱ Turbo>Primus UItra2>Roch Modular PPI(P>0.05)。结论:在检测结果方面,三种检测方法的差异性并不明显,但是在检测结果稳定性方面,Bio-Rad variantⅡTurbo检测方法的结果稳定性最高。 Objective:To compare the detection results of three full-glycemic hemoglobin A1c detection systems under the same storage conditions,and to further explore the stability of different detection methods.Method:Select 50 medical people who were accepted in our hospital from July 2018 to July 2019,and select 2 blood samples in the morning on an empty stomach in the morning and 2 hours after a meal,and inject EDTA-K2 In the anticoagulation test tube,the three detection methods of Bio-Rad variant Ⅱ Turbo,Primus UItra2 and Roch Modular PPI were subsequently used to compare the differences and stability of the results of the three detection methods.Results:The detection results of the three detection methods were not different(P>0.05);in terms of the stability of the detection results,Bio-Rad variant Ⅱ Turbo>Primus UItra2>Roch Modular PPI(P>0.05).Conclusion:In terms of the test results,the difference between the three test methods is not obvious,but in terms of the stability of the test results,the results of the Bio-Rad variant Ⅱ Turbo test method have the highest stability.
作者 孟颖 林朋 MENG Ying;LIN Peng(Department of Clinical Laboratory,Shenyang 242 Hospital,Liaoning Shenyang 110034;Inspection Department of Health Team of Fushun Branch of Armed Police,Liaoning Fushun 113006)
出处 《中国医疗器械信息》 2020年第11期129-130,共2页 China Medical Device Information
关键词 全血糖化血红蛋白A1c 检测系统 稳定性 total glycated hemoglobin A1c detection system stability
  • 相关文献

参考文献5

二级参考文献32

  • 1周新,涂植光.临床生物化学和生物化学检验[M].第3版.北京:人民卫生出版社,2006:57.
  • 2American Diabetes Association. Diagnosis and classification of diabetes mellitus [ J ]. Diabetes Care, 2010,33 (Suppl 1 ) :S62-S69.
  • 3The Diabetes Control and Complications Trail Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mullites [J]. N Engl J ned,1993,329(14):977-986.
  • 4UK Prospective Diabetes Study (UKPDS)Group. Intensive blood-glucose control with conventional treatment and risk of complications in patients with type 2 diabetes ( UKPDS 33 ) [ J ]. Lancet, 1998,352 (9131) :837-853.
  • 5Little RR, Rohlfing CL, Wiedmeyer HM, et al. The national glycohemoglobin standardization program: a five-year progress report [ J ]. Clin Chem, 2001,47 ( 11 ) :1985-1992.
  • 6Steffes M, Cleary P, Goldstein D, et al. Hemoglobin Ao measurements over nearly two decades: sustaining comparable values throughout the Diabetes Control and Complications Trial and the Epidemiology of Diabetes Interventions and Complications study [ J ]. Clin Chem ,2005,51 (4) :753-758.
  • 7Jones W, Scott J, Leary S, et al. Stability of whole blood at -70 degrees C for measurement of hemoglobin A ( 1 c ) in healthy individuals [ J ]. Clin Chem ,2004,50 (12) :2460-2461.
  • 8Little RR, Rohlfing CL, Tennill AL, et al. Effects of sample storage conditions on glycated hemoglobin measurement : evaluation of five different high performance liquid chromatography methods [ J ]. Diabetes Technol Ther,2007,9( 1 ) :36-42.
  • 9王冬环,张传宝.全血糖化血红蛋白用常规方法测定的稳定性研究[J].中华检验医学杂志,2009,32(10):1178-1182. 被引量:9
  • 10杨昌国,许叶,张抗.精密度评价和方法比较中NCCLS评价方案的应用[J].临床检验杂志,1999,17(1):47-49. 被引量:184

共引文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部