期刊文献+

一种IMS信号校正算法

Emendation Arithmetic for Ion Mobility Spectrometry Signal
下载PDF
导出
摘要 目的降低外界环境对离子迁移率的影响。方法讨论用标校物的IMS信号迁移时间对待测物质的IMS信号迁移时间进行校正的算法,通过对待测物质IMS信号进行修正,来减少外界环境的变化对IMS信号迁移时间的影响,提高探测的准确性,最后通过用空气作为标校物进行试验验证。结果迁移区电压的变化对IMS信号的影响降低了96%,迁移区温度的变化对IMS信号的影响降低了53%。结论该方法有效,为该校正技术在IMS信号处理中的应用奠定了基础。 The paper aims to reduce the influence of external environment on ion mobility. In this paper, the algorithm of correcting the IMS signal migration time of the material to be measured by using the IMS signal migration time of the calibrator was discussed. By modifying the IMS signal of the material to be measured, the influence of the change of the external environment on the migration time of the IMS signal was reduced, and the accuracy of the detection was improved. Finally, the air was used as calibration object for verification. The influence of the voltage change in the migration zone on the IMS signal was reduced by 96%. The influence of the temperature in the migration zone on the IMS signal was reduced by 53%. This method is effective, which lays a foundation for the application of this correction technology in IMS signal processing.
作者 郑小保 徐威 张国栋 ZHENG Xiao-bao;XU Wei;ZHANG Guo-dong(The 22nd Research Institute of CETC,Qingdao 266107,China;Shandong Branch of China Mobile Communication Group Design Institute Co.Ltd,Jinan 250001,China)
出处 《装备环境工程》 CAS 2020年第7期12-15,共4页 Equipment Environmental Engineering
关键词 离子迁移谱 迁移率 校正 ion mobility spectrometry mobility rate emendation
  • 相关文献

参考文献9

二级参考文献154

共引文献76

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部