期刊文献+

基于形变补偿的非接触式液体流量算法研究

Research on Contactless Liquid Flow Algorithm Based on Deformation Compensation
下载PDF
导出
摘要 气压驱动式自动腹膜透析机多采用基于气体压强的非接触液体流量算法,文章针对其在治疗过程中流量控制精度受腹腔与透析机相对高度的影响较大的问题,对液体流量计算过程进行分析,采用基于形变补偿的算法,较好地减少了患者腹腔和机器的相对高度变化对流量控制精度的影响,经实验验证,采用形变补偿的算法计算可提升自动腹膜透析机的流量控制精度. Type automatic peritoneal dialysis machine pneumatic drive is used more based on the non-contact liquid gas pressure flow algorithm,in this paper,the flow control accuracy in the process of treatment of the abdominal cavity and a dialysis machine greatly influenced by the relative height of the problem,analyze the liquid flow calculation process,the algorithm based on deformation compensation,better reduce the patients with abdominal cavity and the change of relative height of the machine on the flow control accuracy and the effect of the experimental verification,the deformation compensation algorithm for computing can improve the flow control accuracy of automatic peritoneal dialysis machine.
作者 陈大鹏 丁磊 郝玉哲 CHEN Dapeng;DING Lei;HAO Yuzhe(Jiangsu Automation Research Institute,Lianyungang 222061,China;Linhong Water Conservancy Management Office of Lianyungang,Lianyungang 222061,China)
出处 《现代信息科技》 2020年第11期48-50,共3页 Modern Information Technology
关键词 自动腹膜透析机 流量算法 腹腔高度 形变补偿 automated peritoneal dialysis machine flow algorithm peritoneal height deformation compensation
  • 相关文献

参考文献5

二级参考文献32

  • 1殷明华,陈马连,孔凡让,高立新.LabVIEW在液位监测系统中的应用[J].自动化仪表,2005,26(4):56-58. 被引量:4
  • 2董湘,邹国奎.基于LabVIEW的远程测控方法研究[J].自动化仪表,2006,27(1):6-8. 被引量:26
  • 3刘林,邱继红,王瑞芳.基于Labview开发统计过程控制软件[J].计算机工程与设计,2007,28(2):459-461. 被引量:9
  • 4Burkart J. The future of peritoneal dialysis in the United States : optimizing its use [ J 1. Clinical journal of the American Society of Neph- rology,2009,4( Suppl 1 ) :S125 - S131.
  • 5Yalavarthy R, Teitelbaum I. Peritoneal Dialysis Adequacy:Not Just Small-Solute Clearance[ J ]. Advances in Peritoneal Dialysis, 2008, 24:99 - 103.
  • 6National Kidney Foundation Dialysis Outcomes Quality Initiative:NKF-K/DOQI clinical practice guidelines for peritoneal dialysis adequacy: update 2000 [ J ]. American Journal of Kidney Disease, 2000,37(Suppl 1 ) :S65 - S136.
  • 7Mexican Nephrology Collaborative Study Group. Effects of increased peritoneal clearances on mortality rates in peritoneal dialysis:ADE- MEX, a prospective,randomized, controlled trial [ J ]. Journal of the American Society of Nephrology ,2002,13 (5) : 1307 - 1320.
  • 8Korevaar JC, Boeschoten EW, Dekker FW, et al. What have we learned form NECOSAD? Practical implications for peritoneal dialysis patients [ J ]. Peritoneal dialysis international ,2007,27 ( 1 ) : 11 - 15.
  • 9Lo WK,Ho YW,Li CS,et al. Effect of Kt/V on survival and clinical outcome in CAPD patients in a randomized prospective study [ J ]. Kidney International,2003,64(2) :649 -656.
  • 10Bargman JM, Thorpe KE, Churchill DN , et al. Relative contribution of residual renal function and peritoneal clearance to adequacy of dialysis:a reanalysis of the CANUSA study[ J]. Journal of the Ameri- can Society of Nepbrology,2001,12 (10) :2158 - 2162.