摘要
为了增加过氧化氢低温等离子灭菌器的灭菌效果,提高临床腹腔镜手术器械的灭菌程度,增加灭菌器的使用率,结合强生STERRAD 100s等离子灭菌器的常规维护,阐述3例灭菌器的故障维修实例,供读者参考。
In order to increase the sterilization effect of hydrogen peroxide low-temperature plasma sterilizer,improve the sterilization degree of clinical laparoscopic surgical instruments,and increase the utilization rate of sterilizer,this paper combined with the routine maintenance of Johnson&Johnson sterlad 100s plasma sterilizer,and elaborated three examples of troubleshooting of sterilizer for readers’reference.
作者
杜昱铿
刘月月
DU Yu-keng;LIU Yue-yue(Department of Equipment,Hunan Children’s Hospital,Hunan Changsha 410007;Department of Urology,Hunan Children’s Hospital,Hunan Changsha 410007)
出处
《中国医疗器械信息》
2020年第15期178-179,共2页
China Medical Device Information
关键词
过氧化氢
低温等离子灭菌器
故障处理
维护
hydrogen peroxide
low temperature plasma sterilizer
troubleshooting
maintenance