医用加速器偏转线圈铜导线结垢过程研究

Study on scaling process of copper wire in deflection coil of medical accelerator
下载PDF
导出
摘要 目的:探究医用加速器偏转线圈铜导线结垢过程,分析偏转线圈铜导线堵塞的原因。方法:选用T2黄铜在高纯除盐水中进行腐蚀结垢试验,通过伏安极化、恒电位还原和电子扫描电镜分析T2黄铜垢层的产生过程、化学成分及微观形貌,通过试验研究CO2在水中溶解对T2黄铜垢层剥落的影响。结果:T2黄铜在高纯除盐水中发生氧化还原反应,先生成紫红色的、结构致密的Cu2O内垢层,后生成黑褐色的、结构疏松的CuO外垢层。空气中CO2在水中溶解会加速结构疏松的CuO外垢层溶解剥落。结论:铜导线的氧化是形成铜垢的根本原因,冷却水中CO2的侵蚀会加速铜垢的剥落,从而引发偏转线圈堵塞、过热。 Objective To explore the scaling process of copper wire in the deflection coil of the medical accelerator,and to analyze the blockage cause of deflection coil copper wire.Methods The corrosion test of T2 brass in pure water was carried out in the laboratory.The chemical composition,growth process and microscopic shape of T2 brass scale layer were analyzed by electrochemical voltammetry,potentiostatic reduction and scanning electron microscopy.The effect of CO2 dissolution on the flaking of the copper scale was studied.Results T2 brass reacted with the dissolved oxygen in water to form a purple-red Cu2O inner scale layer and a taupe CuO outer scale layer.The dissolution of CO2 in the air lead to the dissolution and spalling of the loose CuO layer.Conclusion The copper scale is mainly caused by the oxidation of the copper wire,and the erosion of CO2 in the cooling water accelerates the flaking of copper scale,which makes the deflection coilclog and overheat.
作者 李丽 查方林 LI Li;ZHA Fang-lin(Hunan Provincial Maternal and Child Health Care Hospital,Changsha 410005,China;Hunan Xiangdian Test Research Institute Co.,Ltd.,Changsha 410007,China)
出处 《医疗卫生装备》 CAS 2020年第9期76-79,共4页 Chinese Medical Equipment Journal
关键词 医用加速器 偏转线圈 空芯铜导线 T2黄铜 medical accelerator deflection coil hollow core copper wire T2 brass
  • 相关文献

参考文献11

二级参考文献62

共引文献65

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部