高能(450kV)X射线参考辐射场附加过滤研究 被引量:2

Study on Additional Filtration of High Energy(450kV)X-ray Reference Radiation Field
下载PDF
导出
摘要 利用MCNP程序对450kV X光机建立模型,计算各参考辐射质下的原级谱、平均光子能量。对ISO4037-1中已有范围的管电压与附加过滤片厚度关系曲线进行二次拟合,确定模拟厚度范围。利用MCNP计算各附加过滤组合下的平均光子能量值,并与外推法得出的平均光子能量值进行对比,将相对偏差最小时的附加过滤作为最终确定值。从部分能量点的模拟与测量结果的对比可看出,模拟得出的平均光子能量在2%以内与PTB及中国计量院测量结果保持一致,附加过滤最大相对偏差为18%。 The model of 450kV X-ray tube is established by MCNP code,and the primary X-ray spectrums and mean energy of each reference radiation qualities are calculated.The curve of the relationship between the existing range of tube voltage and the thickness of the additional filtration in ISO 4037-1 is fitted quadratic function to determine the thickness range required for simulation.MCNP code is used to calculate the mean photon energy of each additional filtration combination,and compared with the mean photon energy with extrapolation method.The additional filtration with the smallest relative deviation is taken as the final determined value.From the comparison of simulation and measurement results of some energy points,it can be seen that the mean photon energy obtained by simulation is within 2%,which is consistent with the measurement results of PTB and the National Institute of Metrology,and the maximum relative deviation is 18%.
作者 徐阳 高飞 赵瑞 王波 林敏 倪宁 XU Yang;GAO Fei;ZHAO Rui;WANG Bo;LIN Min;NI Ning(China Institute of Atomic Energy,National Key Laboratory for Metrology and Calibwation Techniques,Beijing 102413,China;National Institute of Metrology,Beijing 100029,China)
出处 《宇航计测技术》 CSCD 2020年第3期69-74,共6页 Journal of Astronautic Metrology and Measurement
关键词 X射线 参考辐射场 附加过滤 蒙特卡罗模拟 X-ray Reference radiation filed Additional filter Monte Carlo simulation
  • 相关文献

参考文献4

二级参考文献29

  • 1吴剑,刘荣,王宣银.工业CT研究发展状况及其在流体传动及控制中的应用分析[J].机床与液压,2004,32(10):7-8. 被引量:1
  • 2先武,李时光,王珏.最佳无损检测手段──工业CT技术的发展[J].光电工程,1995,22(4):51-58. 被引量:36
  • 3Krumm M,Kasperl S,Franz M.Reducing non-lindear artifacts of multi-material objects in industrial 3D computed tomography [J].NDTand E International,2008,41(4):242-251.
  • 4Bartscher M,Hilpert U? Goebbels J,et ai.Enhancement and proof of accuracy of industrial computer tomography(CT)measurements[J].Annals of the CIRP,2007,56(1):495-498.
  • 5Nalcioglu O,Lou R Y.Post-reconstruction method for beam hardening in computerized tomography[J].Phys Med Biol,1979,24(2):33-40.
  • 6Van de Casteele E,Van Dyck D,Sijbers J,et al.A model-based correction method for beam hardening artefacts in X-ray microtomography[J].J ournal of X-ray Science and Technology,2004,12(1):43-57.
  • 7Hammersberg P,Mangard M.Correction for beam hardening artefacts in computerised tomography [J].Journal of X-ray Science andTechnology,1998,8(1):75-93.
  • 8Davis G W,Wallenslager S T.Improvement of chest region CT images through automated gray-level remapping[J].IEEE Trans on Medi-cal imaging,1983,5(1):30-34.
  • 9Rattarangsi A,Chin R T.Scale-based detection of corner of planar curves[J].IEEE Trans on Pattern Anal Mach Intell,1992,14(2):430-449.
  • 10Gonzalez M,Dominguez G,Bathias C.Some results from X-ray computed tomography applied to metal matrix composites[J],Journal ofComposites Technology &-Research,2000,22(1):45-48.

共引文献17

同被引文献13

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部