期刊文献+

液相色谱质谱法快速测定水中痕量涕灭威和克百威 被引量:5

Rapid determination of trace aldicarb and carbofuran in water by HPLC-MS/MS
下载PDF
导出
摘要 建立了快速检测水中痕量涕灭威和克百威的高效液相色谱-质谱方法。水样经C18柱分离,在电喷雾离子源正离子模式下,利用动态多反应监测模式采集数据,采用内标法进行定量分析。结果表明,涕灭威和克百威分别在0.020~0.640、0.005~0.160μg/L范围内线性关系良好(r≥0.998),方法检出限分别为0.02、0.005μg/L。低、中、高3个加标浓度水平下,加标回收率为90.4%~102%,相对标准偏差为2.2%~18.4%。 A rapid determination method of trace aldicarb and carbofuran in water by high performance liquid chro-matography-mass spectrometry was established.Water samples were separated by C18 column.The data was collected and analyzed with dynamic multiple reaction monitoring under ESI positive ion mode.The internal standard method was used for quotatives analysis.The results showed that aldicarb and carbofuran had good linearity(r≥0.998),in the range of 0.020-0.640μg/L and 0.005-0.160μg/L respectively.The detection limits of aldicarb and carbofuran were 0.02μg/L and 0.005μg/L respectively.The recoveries were 90.4%-102%and RSD were 2.2%-18.4%at low,medium and high spiked concentration.
作者 李利荣 张静 刘殿甲 吴宇峰 张肇元 王效国 张海波 许亮 Li Lirong;Zhang Jing;Liu Dianjia;Wu Yufeng;Zhang Zhaoyuan;Wang Xiaoguo;Zhang Haibo;Xu Liang(Tianjin Eco-Environmental Monitoring Center,Tianjin 300190,China;Tianjin Tianbinruicheng Environmental Technology and Engineering Co.,Ltd.,Tianjin 300190,China)
出处 《工业水处理》 CAS CSCD 北大核心 2020年第10期111-115,共5页 Industrial Water Treatment
关键词 涕灭威 克百威 高效液相色谱 质谱法 快速测定 aldicarb carbofuran HPLC MS/MS rapid determination
  • 相关文献

参考文献5

二级参考文献50

共引文献59

同被引文献51

引证文献5

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部