期刊文献+

激光光散射-凝胶渗透色谱联用系统的校准 被引量:1

Calibration of laser light scattering-gel permeation chromatography
下载PDF
导出
摘要 建立了激光光散射-凝胶渗透色谱联用系统的校准方法。色谱柱为HT3+HT4型凝胶渗透色谱柱,柱温为25℃,流动相为四氢呋喃,流量为1.0 mL/min,采用聚苯乙烯作为标准样品进行试验。测定结果相对偏差的绝对值不大于10%,重复性(相对标准偏差)小于5%(n=6)。当聚苯乙烯相对分子质量为177670时,不确定度为10600(k=2)。该方法操作简便、快速,可用于激光光散射仪与凝胶渗透色谱联用系统的校准。 The calibration method of laser light scattering-gel permeation chromatography was established.The chromatographic column was type of HT3+HT4 gel permeation chromatographic column,the column temperature was 25℃,the flow solvent was THF with the rate of 1.0 mL/min,and polystyrene was as the standard sample.The relative error was less than±10%,the relative standard deviation of the detection results was less than 5%(n=6).When the relative molecular mass of polystyrene was 177670,the uncertainty was 100660(k=2).The method is rapid,and it is suitable for calibration of laser light scattering-gel permeation chromatography,achieving the tracing of measuring parameter to the social benchmarks.
作者 王佳雯 张翠珍 于科青 苏伟伟 温泽强 李阳 贾巧连 朱麟鹏 WANG Jiawen;ZHANG Cuizhen;YU Keqing;SU Weiwei;WN Zeqiang;LI Yang;JIA Qiaolian;ZHU Linpeng(Inner Mongolia Institute of Aerospace Machinery Testing,Hohhot 010010,China;Inner Mongolia Synthetic Chemical Engineering Institute,Hohhot 010010,China)
出处 《化学分析计量》 CAS 2020年第S01期86-91,共6页 Chemical Analysis And Meterage
关键词 激光光散射-凝胶渗透色谱联用系统 校准 聚苯乙烯 laser light scattering-gel permeation chromatography combined system calibration polystyrene
  • 相关文献

参考文献11

二级参考文献60

共引文献40

同被引文献7

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部