摘要
通过了解PB840型呼吸机的基本组成与呼吸模式,对电磁阀裂缝漏气所导致的潮气量误差过大、氧电池使用时间过长所导致的氧浓度偏低以及空气流量传感器污染所导致的潮气量不准3种故障的发生原因进行分析,提出故障排除方法。医院临床工程师应重视对流量传感器的维护,加强日常巡检和呼吸机定期自检工作,以确保医疗设备的安全使用。
By understanding the basic composition and respiratory pattern of PB840 ventilator,the occurrence reasons of three kinds of faults,which included excessive tidal volume error caused by the leakage of the crevice of solenoid valve,lower oxygen concentration caused by too long use time of oxygen battery and inaccurate tidal volume caused by the pollution of air flow sensor,were analyzed,and the relevant elimination methods of faults were proposed.Clinical engineer of hospital should pay attention to the maintenance of flow sensor and strengthen routine routing inspection and termly self-inspection for ventilator so as to ensure the safe use of medical equipment.
作者
孙哲
田金
许锋
SUN Zhe;TIAN Jin;XU Feng(Department of Medical Engineering,Peking University Third Hospital,Beijing 100191,China)
出处
《中国医学装备》
2020年第10期224-227,共4页
China Medical Equipment
关键词
呼吸机
呼吸模式
流量传感器
氧电池
潮气量
Ventilator
Ventilation mode
Flow sensor
Oxygen battery
Tidal volume