摘要 目的设计基于SMR170体模磁共振成像系统质量控制检测的线性度自动测量程序,实现线性度参数的自动测算。方法在Matlab平台设计编写了图像处理算法,对线性度参数进行自动测量,并采用Bland-Altman统计学方法对手工测量和自动测量两种方法的一致性进行了分析。结果利用Matlab平台设计编写的自动检测方法可实现线性度的自动检测,并且与手工测量方法有较好的一致性。结论线性度自动检测程序与手工测量可相互替代。 Objective To design an automatic linearity measurement program based on the quality control test of the SMR170 phantom MRI system in order to realize the automatic measurement of linearity parameters.Methods The image processing algorithm was designed and written on the Matlab platform,and the linearity parameters were automatically measured.The consistency between manual measurement method and automatic measurement method was analyzed by Bland-Altman statistical method.Results The automatic measurement program based on Matlab platform could measure linearity automatically,which had good consistency with manual measurement method.Conclusion The automatic linearity detection program and manual measurement can be substituted for each other.
作者 梁永刚 付丽媛 陈坚 黄志峰 熊晖 李良杰 钟群 肖慧 许尚文 陈自谦 LIANG Yonggang;FU Liyuan;CHEN Jian;HUANG Zhifeng;XIONG Hui;LI Liangjie;ZHONG Qun;XIAO Hui;XU Shangwen;CHEN Ziqian(Department of Radiology,900th Hospital of Joint Logistics Support Force(Formerly Fuzhou General Hospital of Nanjing Military Region),Fuzhou Fujian 350025,China;Department of Radiology,The Affiliated Dongfang Hospital of Xiamen University,Fuzhou Fujian 350025,China)
出处 《中国医疗设备》 2020年第11期12-15,共4页 China Medical Devices
基金 国家重点研发计划数字诊疗装备研发重点专项子课题(2016YFC0103103) 军区重大课题(14ZX23) 医院科研计划(2019Q02)。
关键词 磁共振成像系统 SMR170体模 质量控制检测 线性度 自动测量 magnetic resonance imaging system SMR170 phantom quality control test linearity automatic measurement
  • 相关文献

参考文献14

二级参考文献105

共引文献250

同被引文献24

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部