摘要
从密度计(浮计型)的检定全过程进行分析,对检定采用的标准器、配套设备、毛细常数修正方法、温度的控制、操作细节等均进行细致地解析并量化误差,针对每项影响因素提出合理的正确操作,为提高密度计检定准确性提供了科学的操作建议。
This paper analyzes and quantifies the errors involved in the whole verification process of densimeter(float type),including standard apparatus,supporting equipment,capillary constant correction method,temperature control,operation details,and so on.By putting forward reasonable and correct operation for each influencing factor,this paper proposes scientific operation suggestions for improving the accuracy of densimeter verification.
作者
郭丽宏
GUO Lihong(Laboratory and Measurement Center of Sinopec Tianjin Branch,Tianjin 300271,China)
出处
《计量科学与技术》
2020年第9期13-17,共5页
Metrology Science and Technology
关键词
密度计
检定
标准器
误差
densimeter
verification
standard apparatus
error