摘要
分析Aquilion ONE 640层CT出现的故障现象,依据先硬件后软件的故障排查原则,对故障设备进行检测,通过更换扫描序列控制(SS/ADI)模块解决"TaskStartererror"故障。SS/ADI模块故障应以设备启动时文档成功下载为判断依据,可为类似的设备故障维修提供参考。
To analyze fault phenomenon of Aquilion ONE 640 slices computed tomography(CT),and propose troubleshooting principle that software was behind hardware.After the fault equipment was detected,the fault of"TaskStarter error"was resolved through replaced scanning sequence(SS/ADI)module.The success downloading of documents when equipment was launched should be used as basis of determination of judging SS/ADI fault,and it can provide references for the maintenance of similar equipment faults.
作者
姜宏涛
孔明
陈燚
王明刚
JIANG Hong-tao;KONG Ming;CHEN Yi(Department of Medical Engineering,People's Liberation Army Hospital 971,QingDao,266071,China;不详)