摘要
简要介绍了光学相干断层扫描仪的工作原理,重点阐述了RTVue XR光学相干断层扫描仪在使用中出现的开机异常、左右眼不识别、软硬件连接错误等故障维修过程,为维修人员维修类似故障提供了参考。最后指出了平时应注意维护保养才能降低设备故障率,保证设备的正常运行。
The working principle of RTVue XR Avanti Ophthalmic Optical Coherence Tomography System was briefly described.The treatment of various faults of the system were introduced in detail,involving in abnormal start-up,non-identification of left and right eyes and wrong connection of software and hardware.References were provided for the maintenance personnel when treating similar faults.It's pointed out daily maintenance contributed to decreasing fault rate and ensuring normal running of the system.
作者
惠博阳
潘登
李爱萍
HUI Bo-yang;PAN Deng;LI Ai-ping(Tianjin Ophthalmological Hospital,Tianjin 300020,China)
出处
《医疗卫生装备》
CAS
2020年第12期107-108,共2页
Chinese Medical Equipment Journal
关键词
RTVue
XR光学相干断层扫描仪
工作原理
维护保养
故障维修
RTVue XR Avanti Ophthalmic Optical Coherence Tomography System
working principle
maintenance
fault treatment