Accuracy and Easiness of Three Self-Administered Pure Tone Audiometry Apps in a Normal Hearing Population
下载PDF
导出
摘要 目的评估基于移动智能终端的自助测听应用程序(App)在听力正常人群中的准确度和便捷性。方法在基于iOS系统的移动智能终端上,自行下载NIOSH SLM(Sound Level Meter,声级计)App及三款自助纯音测听App(*伴/Mimi/*通)。应用NIOSH SLM选定本底噪声小于40dB A的安静环境,招募20名听力正常的受试者,分别自助应用三款测听App完成双耳0.25、0.5、1、2、4、8kHz六个倍频程纯音的气导听阈测试;同时用秒表记录各应用程序测听的耗时。于医院听力中心完成入组筛查(鼓膜完整、外耳道无耵聍堵塞、A型鼓室图等)后,采用Hughson-Westlake法以GSI 61诊断型听力计对受试者进行气导纯音测听,并用秒表记录耗时。比较自助测听App和临床纯音测听所测气导听阈及耗时的差异。结果三款自助纯音测听App中,*伴、Mimi两者测得的气导听阈与临床纯音测听结果一致性较好,耗时也无显著性差异。*通测得的气导听阈与临床纯音测听在250Hz、500Hz有显著性差异,其他频率无显著差异,但耗时较其他两款应用程序及临床纯音测听显著缩短。结论*伴、Mimi所测气导听阈与临床纯音测听法所获听阈的一致性较好;*通所测气导听阈在低频区有明显偏差,但测听所用耗时较短。现行多款基于移动智能终端的自助测听App,若要获得快捷可靠的自助纯音气导测听结果,仍有很大的改进空间。 Objective To evaluate the accuracy and easiness of mobile device-based self-service audiometric applications(apps)in a normal hearing population.Methods The NIOSH SLM(Sound Level Meter)app and three audiometric apps(*Assistant/Mimi/*Tone)were downloaded onto iOS smartphones.The NIOSH SLM application was used in a quiet environment with background noise levels less than 40 dBA.Twenty subjects with normal hearing were recruited to complete binaural air conduction thresholds test at six octave frequencies(0.25kHz,0.5 kHz,1 kHz,2 kHz,kHz 4 and 8 kHz)using the three audiometric apps and the time required to complete testing using each app was recorded.The same subjects were also screened(eardrum intact,external ear cerumen unblocked,type A tympanogram,etc.)and received formal air-conduction pure-tone audiometry using a GSI 61 diagnostic audiometer and the Hughson-Westlake method at the hospital hearing center,with the test time also recorded.Differences in hearing thresholds and test time between audiometric apps and formal audiometry were compared.Results Among the three pure-tone self-service audiometric apps,thresholds measured by the*Assistant and Mimi apps were highly consistent with those obtained via formal pure-tone audiometry,with no significant difference in test time.Thresholds measured by the*Tone apps were different from those by formal audiometry at 250 Hz and 500 Hz,although with significantly less amount of time needed than the other two apps and formal audiometry.Conclusion Air conduction hearing thresholds measured by the*Assistant and Mimi apps are in good agreement with clinical pure tone audiometry;but those measured by the*Tone app show significant deviations at low frequencies.The*Tone app requires the shortest test time.Much remains to be improved for many of the existing mobile device-based self-assessment audiometric apps to provide fast and reliable selfservice pure tone audiometric screening.
作者 周鹏 李赟硕 周颖 郗昕 ZHOU Peng;LI Yunshuo;ZHOU Ying;XI Xin(Department of Otolaryngology,Union Hospital of Tongji Medical College,Huazhong University of Science and Technology,Wuhan 430022 China;Department of Otolaryngology,Beijing Tsinghua Changgung Hospital Affiliated to Tsinghua University,Beijing 102218 China;College of Otolaryngology Head and Neck Surgery,Chinese PLA General Hospital,National Clinical Research Center for Otolaryngologic Diseases,Key Lab of Hearing Impairment Science of Ministry of Education,Key Lab of Hearing Impairment Prevention and Treatment of Beijing,Beijing,100853 China)
出处 《中华耳科学杂志》 CSCD 北大核心 2020年第6期1038-1044,共7页 Chinese Journal of Otology
基金 高性能及免验配助听器临床评估与推广应用示范(2020YFC2004005)。
关键词 纯音测听 自助测听 应用程序 Pure Tone Audiometry Self-Assessment Hearing Test Applications
  • 相关文献

参考文献6

二级参考文献29

  • 1陈洪文,于黎明.听力计检定和声场校准[J].中华耳科学杂志,2003,1(4):63-67. 被引量:18
  • 2王洪田.耳鸣的诊断治疗新进展[J].实用医学杂志,2005,21(2):114-116. 被引量:96
  • 3国家技术监督局.声学测听方法纯音气导和骨导听阈基本测听法.GB/T 16403-1996.
  • 4Imamura S, Nozawa I, Imamura M, et al. Clinical observations on acute low-tone sensorineural heating loss. Survey and analysis of 137 patients. Ann Otol Rhinol Laryngol 1997; 106(9):746-750.
  • 5American Speech-Language-Hearing Association. Guidelines for de- termining threshold level for speech. ASHA, 1988, 30, 85-89.
  • 6BrannstrSm K J, Grenner J. Hypobaric pressure exposure effects on cochlear frequency selectivity in fluctuating, low-frequency hearing loss. J Laryngol Otol. 2009, 123(7): 710-717.
  • 7Brannstrom KJ, Grenner J. Long-term measurement of binaural in- tensity and pitch matches. I. Normal hearing. Int J Audiol. 2008, 47: 59-66.
  • 8Margolis RH, Saly GL, Le C, et al. Qualind: A method for assessing the accuracy of automated tests. J Am Acad Audioh2007, 18(1): 78-89.
  • 9Stach BA, Davis-Thaxton ML, Jerger J.Improving the efficiency of speech audiometry: computer-based approach.J Am Acad Audiol. 1995, 6(4):330-333.
  • 10Letz R. Continuing challenges for computer-based neuropsychologi- cal tests. Neurotoxicology.2003, 24 (4-5):479-489.

共引文献13

同被引文献49

引证文献6

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部