期刊文献+

漳州龙师井水氡观测恒温脱气方法实验 被引量:1

Experiment on the application of thermostatic deaeration method in well water radon observation in Longshi of Zhangzhou
下载PDF
导出
摘要 利用恒温水浴仪可以控制水氡样品的温度保持恒定,在恒温条件下进行脱气观测,可以保证水氡测值的稳定性。为验证该方法是否适用于漳州龙师井(水温年均值97℃)水氡观测,将同一水样,利用恒温水浴仪,将水温分别控制在20.0℃、22.5℃、25.0℃、27.5℃、30.0℃,进行一系列比测实验,结果表明:在25.0℃—27.5℃的恒温条件下,氡浓度测值更为稳定,变化更小,可进一步提高水氡观测的质量和稳定性,说明在龙师井采用恒温脱气方法是可行的。 Constant temperature water bath instrument can keep the temperature of water radon sample constant.Deaeration under a constant temperature condition can guarantee the stability of the measurement of water radon.To verify whether this method is suitable for the water radon observation in Longshijing,Zhangzhou(annually average water temperature is 97℃),we take the same water sample and utilize the constant temperature water bath instrument to keep the water temperature at 20.0℃,22.5℃,25.0℃,27.5℃and 30.0℃,respectively to carry out a series of contrast test experiments.The results show that measured radon potency is more stable under 25.0℃—27.5℃constant temperature conditions with smaller changes and is able to further improve the quality and stability of water radon observation,indicating that it is feasible to adopt the constant temperature degas method in Longshijing.
作者 朱继承 林国元 郑辰禾 王晓韡 邓聪 ZHU Jicheng;LIN Guoyuan;ZHENG Chenhe;WANG Xiaowei;DENG Cong(Fujian Earthquake Agency,Fuzhou 350003,China)
机构地区 福建省地震局
出处 《地震地磁观测与研究》 2020年第5期118-121,共4页 Seismological and Geomagnetic Observation and Research
基金 福建省地震局台站基金——福建水氡恒温脱气观测的实验研究(项目编号:T201409)。
关键词 水氡观测 恒温脱气 水温 氡浓度 比测实验 water radon observation thermostatic deaeration water temperature radon concentration experiment analysis
  • 相关文献

参考文献5

二级参考文献13

共引文献14

同被引文献5

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部