摘要 以PpIX光敏剂的生物组织荧光光谱定量检测为目的,采用蒙特卡罗(MC)方法模拟不同组织,的荧光光谱和漫反射光谱,并给出荧光光谱和漫反射光谱的校准方法,通过自主搭建组织荧光光纤检测系统,分别对3种组织光学模拟液中的光敏剂荧光进行实验测试和光谱校准,得出3种组织光学模拟液的原始荧光光谱离散系数为28%,经过构建的算法校准后,离散系数减小到3%.结果表明此方法能有效提高组织模拟液的荧光光谱测量准确性. Monte Carlo(MC) method was used to simulate the fluorescence spectrum and diffuse reflectance spectrum of different tissues,and a calibration method of fluorescence spectrum and diffuse reflection spectrum was proposed.The fluorescence of photosensitizers in three kinds of tissue optical simulators were tested and calibrated by self-developed tissue fluorescence optical fiber detection system.The results showed that the dispersion coefficient of the original fluorescence spectrum of the three tissue optical simulators was 28%.After the calibration of the algorithm constructed in this paper,the dispersion coefficient was reduced to 3%.The results show that this method can effectively improve the accuracy of fluorescence spectrum measurement of tissue simulation solution.
作者 邹健 谢树森 孟娜 黄正 ZOU Jian;XIE Shusen;MENG Na;HUANG Zheng(Key Laboratory of OptoElectronic Science and Technology for Medicine of Ministry of Education,College of Photonic and Electronic Engineering,Fujian Normal University,Fuzhou 350007,China)
出处 《福建师范大学学报(自然科学版)》 CAS 2021年第1期57-61,共5页 Journal of Fujian Normal University:Natural Science Edition
基金 国家自然科学基金面上项目(61675044)。
关键词 光敏剂 组织荧光 荧光光谱 光谱校准 photosensitizer tissue fluorescence fluorescence spectrum spectral calibration
  • 相关文献

参考文献1

二级参考文献20

  • 1kKanick S C, Davis S C, Zhao Y, et al: Dual-channel red/blue fluorescence dosimetry with broadband reflectance spectroscopic correction measures protoporphyrin IX production during photodynamic therapy of actinic keratosis [J]. Journal of Biomedical Optics, 2014, 19(7) : 075002.
  • 2Bradley R S, Thorniley M S. A review of attenuation correction techniques for tissue fluorescence[J]. Journal of the Royal Society Interface, 2006, 3(6): 1-13.
  • 3Brooks S, Hoy C L, Amelink A, et al: Sources of variability in the quantification of tissue optical properties by multidiameter single-fiber reflectance and fluorescence spectroscopy [J] . Journal of Biomedical Optics, 2015, 20 (5): 057002.
  • 4Liu C B, Rajaram N, Vishwanath K, et al. Experimental validation of an inverse fluorescence Monte Carlo model to extract concentrations of metabolically relevant fluorophores from turbid phantoms and a murine tumor model[J]. Journal of Biomedical Optics, 2012, 17(7) : 077012.
  • 5Vald6s S P, Leblond F, Kim A, et al: A spectrally constrained dual-band normalization technique for protoporphyrin IX quantification in fluorescence-guided surgery[J]. Optics Letters, 2012, 37(11): 1817-1819.
  • 6Kaniek S C, Robinson D J, Sterenborg H J, et al: Extraction of intrinsic fluorescence from single fiber fluorescence measurements on a turbid medium[J]. Optics Letters, 2012, 37(5) : 948-950.
  • 7Jacques S L. Optical properties of biological tissues: A review[J]. Physics in Medicine and Biology, 2013, 58(11): R37- R61.
  • 8Poli R, Kennedy J, Blackwell T. Particle swarm optimization[J]. Swarm Intelligence, 2007, 1(1): 33-57.
  • 9Kennedy J. Particle swarm optimization[M]. //Encyclopedia of machine learning. Heidelberg: Springer, 2010:760-766.
  • 10Hull E, Ediger M, Unione A, et al: Noninvasive, optical detection of diabetes: Model studies with porcine skin[J] Optics Express, 2004, 12(19): 4496-4510.