期刊文献+

我国心肌标志物检验项目性能规范的设定 被引量:5

Establishment of performance specifications for myocardial markers in China
下载PDF
导出
摘要 目的制定我国心肌标志物———心肌肌钙蛋白I(cTnI)、心肌肌钙蛋白T(cTnT)、肌红蛋白(Myo)、同型半胱氨酸(Hcy)的性能规范,包括允许总误差(TEa)、允许不精密度(Ia)和允许偏倚(Ba)。方法根据米兰会议性能规范设定模式中的模式二(基于生物学变异)和模式三(基于当前技术水平)设定性能规范。结果基于生物学变异推荐cTnI的TEa、Ia和Ba分别为27.91%(适当)、10.54%(最低)、10.52%;基于生物学变异推荐cTnT的TEa和Ia分别为24.47%(最佳)、15.30%(适当);基于生物学变异推荐Myo的TEa、Ia和Ba分别为19.60%(适当)、10.43%(最低)、2.39%;基于当前技术水平推荐Hcy的TEa、Ia和Ba分别为20.00%、8.50%、5.98%。结论制定了符合我国临床实验室当前检测水平的心肌标志物性能规范。 Objective To establish the performance specifications of myocardial markers such as cardiac troponin I(cTnI),cardiac troponin T(cTnT),myoglobin(Myo)and homocysteine(Hcy)in China,including allowable total error(TEa),allowable imprecision(Ia)and allowable bias(Ba).Methods The performance specifications were set based on biological variation(BV)(Model 2)and the state of the art(Model 3)defined in the Milan Conference.Results The appropriate TEa,Ia and Ba of cTnI set by BV were 27.91%(desirable),10.54%(minimum)and 10.52%,respectively.The appropriate TEa and Ia of cTnT set by BV were 24.47%(optimal)and 15.30%(desirable),respectively.The appropriate TEa,Ia and Ba of Myo set by BV were 19.60%(desirable),10.43%(minimum)and 2.39%,respectively.The appropriate TEa,Ia and Ba of Hcy set by the state of the art were 20.00%,8.50%and 5.98%,respectively.Conclusion The performance specifications of myocardial markers conformed to the current detection level of clinical laboritories in China are established successfully.
作者 杜雨轩 张诗诗 张传宝 汪静 王薇 王治国 DU Yuxuan;ZHANG Shishi;ZHANG Chuanbao;WANG Jing;WANG Wei;WANG Zhiguo(National Center for Clinical Laboratories/Beijing Engineering Research Center of Laboratory Medicine,Beijing Hospital,National Center of Gerontology,Institute of Geriatric Medicine,Chinese Academy of Medical Sciences,Beijing 100730,China)
出处 《临床检验杂志》 CAS 2021年第1期47-52,共6页 Chinese Journal of Clinical Laboratory Science
基金 国家自然科学基金(81871737)。
关键词 性能规范 心肌标志物 允许总误差 允许不精密度 允许偏倚 performance specification myocardial marker allowable total error allowable imprecision allowable bias
  • 相关文献

参考文献2

二级参考文献25

  • 1Kallner A, McQueen M, Heuck C. The Stockholm Consensus Conference on quality specifications in laboratory medicine, 25-26 April 1999[J]. Scand J Clin Lab Invest, 1999, 59(7) : 475-476.
  • 2Fraser CG. Biological variation: from principles to practice [ M ]. Washington, DC : AACC Press, 2001 : 29-66.
  • 3Tonks DB. A study of the accuracy and precision of clinical chemistry determinations in 170 Canadian laboratories [ J ]. Clin Chem, 1963,9 (2) : 217-233.
  • 4Barnett RN. Medical significance of laboratory results[ J]. Am J Clin Pathol, 1968,50(6) : 671-676.
  • 5Cotlove E, Harris EK, Williams GZ. Biological and analytic components of variation in long-term studies of serum constituents in normal subjects. III. Physiological and medical implications[J]. Clin Chem, 1970,16(12) :1028-1032.
  • 6Fraser CG, Petersen PH. Desirable standards for laboratory tests if they are to fulfil medical needs[J]. Clin Chem, 1993, 39(7) :1447-1455.
  • 7Fraser CG. Quality specifications in laboratory medicine [ J ]. Ups J Med Sci, 1990,95(3) : 229-232.
  • 8Fraser CG. The 1999 Stockholm Consensus Conference on quality specifications in laboratory medicine [ J ]. Clin Chem Lab Med, 2015,53(6) : 837-840.
  • 9Rej R, editor. 1992 Clinical Chemistry Forum. Accuracy and precision goals in clinical chemistry testing : can they be defined by medical relevance [ J ]. Clin Chem, 1993,39 ( 7 ) : 1446-1544.
  • 10Current status of blood cholesterol measurement in clinical laboratoties in the Unites States : a report from the Laboratory Standardization Panel of the National Cholesterol Education Program [ J ]. Clin Chem, 1988,34( 1 ) : 193-201.

共引文献15

同被引文献23

引证文献5

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部