摘要 研究了一种经皮无线供能系统.首先通过电路分析得到系统电压增益和传输效率等特性函数,同时根据经皮无线供能典型技术参数进行特性分析,在此基础上设计了基于能量注入的变频恒压控制方案,使得无线供能系统在负载和传输距离变化时始终高效率运行.搭建了经皮无线供能的实验系统,实验结果验证了理论分析和设计方案的正确性,在固定传输距离的整个负载变化范围内整机效率基本保持恒定,典型传输距离内的整机效率达到83%以上.利用多物理场仿真软件进行人体组织安全性仿真实验,仿真结果表明最大电场强度、比吸收率和最高温度均低于限值. A wireless transcutaneous energy transfer(TET)system is researched.The characteristic functions of system voltage gain and transmission efficiency are obtained by circuit analysis.Meanwhile,according to the typical technical parameters of the TET system,a characteristic analysis is conducted.Therefore,based on the energy injection technique,a variable frequency constant voltage control method is proposed which enables the TET system to operate at a high efficiency all time when both the load and the transfer distance change.The experimental set of the TET system is arranged.The experimental results have verified the correctness of the theoretical analysis and design scheme.When the transfer distance is fixed,the overall efficiency remains constant in the whole load variation range.The overall efficiency of the TET system is above 83%within typical transfer distances.The multi-physics simulation software is used to simulate the human tissue safety.The simulation results show that the maximum electric field strength,the specific absorption rate,and the maximum temperature are lower than their corresponding limitations.
作者 谢岳 沈鹏飞 蒋晓丽 XIE Yue;SHEN Pengfei;JIANG Xiaoli(Institute of Mechanical and Electrical Engineering,China Jiliang University,Hangzhou 310018,China)
出处 《上海交通大学学报》 EI CAS CSCD 北大核心 2021年第2期196-205,共10页 Journal of Shanghai Jiaotong University
关键词 经皮无线供能 恒压 变频 能量注入 wireless transcutaneous energy transfer(TET) constant voltage variable frequency energy injection
  • 相关文献

参考文献7

二级参考文献86

  • 1MA Guan-ying YAN Guo-zheng HE Xiu.Study on Wireless Power Transmission for Gastrointestinal Microsystems Based on Inductive Coupling[J].Chinese Journal of Biomedical Engineering(English Edition),2007,16(2):71-78. 被引量:8
  • 2Wilkoff B L,Bello D,Taborsky M,et al.Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment[J].Heart Rhythm,2011,8(1):65-73.
  • 3Amin M S,Ellenbogen K A.Focus on management of pacemaker and ICD advisories,recalls,and alerts[J].Current Treatment Options in Cardiovascular Medicine,2006,8(5):347-352.
  • 4Medtronic.New Research Shows Significant Benefits of Medtronic Deep Brain Stimulation Earlier in Progression of Parkinsons Disease[R/OL].[2014-10-23].http://wwwp.medtronic.com/Newsroom/NewsReleaseDetails.do?itemId=1360783948456&lang=en_CA.
  • 5Lovik R D,Abraham J P,Sparrow E M.Potential tissue damage from transcutaneous recharge of neuromodulation implants[J].International Journal of Heat and Mass Transfer,2009,52(15):3518-3524.
  • 6Smith D K,Lovik R D,Sparrow E M,et al.Human tissue temperatures achieved during recharging of new-generation neuromodulation devices[J].International Journal of Heat and Mass Transfer,2010,53(15):3292-3299.
  • 7Lovik R D,Abraham J P,Sparrow E M.Surrogate human tissue temperatures resulting from misalignment of antenna and implant during recharging of a neuromodulation device[J].Neuromodulation:Technology at the Neural Interface,2011,14(6):501-511.
  • 8St.Jude Medical.Letter to Patients from St.Jude Medical:July 2012.[R/OL].[2014-10-22].http://professional.sjm.com/professional/resources/product-performance/eon-eonmini-important-info/physician-communication.
  • 9Kudo H,Ogawa K,Oodachi N,et al.Detection of a metal obstacle in wireless power transfer via magnetic resonance[C]//2011 IEEE 33rd International Telecommunications Energy Conference(INTELEC).Piscataway,NJ:IEEE Press,2011:1-6.
  • 10Yu X,Sandhu S,Beiker S,et al.Wireless power transfer with the presence of metallic planes[J].Applied Physics Letters,2011,99(21),214102.

共引文献71

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部