宝石能谱CT与常规CT在颌面部检查中辐射剂量及图像质量的对比研究 被引量:4

Gemstone Spectral CT Imaging in the Maxillofacial Examination:Initial Results of Radiation Exposure and Image Quality Compared with Conventional CT
下载PDF
导出
摘要 目的探讨宝石能谱CT扫描在颌面部检查中对辐射剂量及图像质量的影响。方法收集颌面部CT平扫患者40名,分成两组(A组+B组),每组20名。A组采用单源瞬时kVp切换能谱CT扫描模式,管电压140、80 kVp瞬时(0.5 ms)切换,重建出单能量70 keV水平图像。B组行常规扫描,管电压120 kVp。两组其他扫描参数相同。对两组图像的信噪比、图像显示质量及辐射剂量进行对比。结果A组和B组间脂肪、气体、腮腺和髁突间的信噪比差异无统计学意义(-15.02±5.130和-15.22±6.105,P=0.867>0.05;-151.71±41.103和-169.77±73.295,P=0.179>0.05;0.39±2.821和1.22±3.694,P=0.263>0.05;11.07±5.652和11.56±5.873,P=0.701>0.05),而咬肌、颌下腺和玻璃体的信噪比差异有统计学意义(8.58±2.637和12.31±3.763,P<0.001;7.17±2.407和10.24±5.147,P=0.001;0.97±0.511和2.91±0.867,P<0.001)。主观评价两组图像质量无统计学差异。A组和B组有效剂量分别为:2.80±0.432 mSv,3.54±0.737 mSv(P=0.001)。结论宝石能谱CT扫描在不影响图像质量和影像诊断的前提下,与常规扫描相比,能有效降低辐射剂量。 Objective To explore the effect of gemstone spectral CT scanning on radiation exposure and image quality in maxillofacial examination.Methods Forty patients with maxillofacial CT plain scan were divided into two groups(A group and B group).Each gourp included 20 patients.In A group,the gemstone spectral imaging(GSI)with single source instantaneous switch of kVp(140 kVp,80 kVp)on 0.5 ms was performed.And single energy 70 keV level image was reconstructed.In B group,the conventional polychromatic imaging(CPI)with the tube voltage 120 kVp was performed.The other scanning parameters of two groups were the same.The signal-to-noise ratio(SNR),image display quality and radiation dose were compared between the two groups.Results There was no significant difference in SNR of fat,gas,parotid gland and condyle between A group and B group(-15.02±5.130 and-15.22±6.105,P=0.867;-151.71±41.103 and-169.77±73.295,P=0.179;0.39±2.821 and 1.22±3.694,P=0.263;11.07±5.652 and 11.56±5.873,P=0.701).There was significant difference in the SNR of masseter,submandibular gland and vitreum between A group and B group(8.58±2.637 and 12.31±3.763,P<0.001;7.17±2.407 and 10.24±5.147,P=0.001;0.97±0.511 and 2.91±0.867,P<0.001).There was no statistical difference in subjective evaluation between two groups.And ED of A group and B group were 2.80±0.432 mSv and 3.54±0.737 mSv(P=0.001).Conclusion GSI can effectively reduce radiation dose compared with CPI without affecting image quality and clinical image diagnosis.
作者 高维青 龚海燕 沈阳 刘玉 GAO Weiqing;GONG Haiyan;SHEN Yang;LIU Yu(Department of Radiology,Ninth People’s Hospital,Shanghai Jiao Tong University School of Medicine,Shanghai 200011,China)
出处 《中国医疗设备》 2021年第3期77-80,共4页 China Medical Devices
关键词 宝石能谱 体层摄影 X线计算机 颌面部 辐射剂量 图像质量 gemstone spectrum tomography X-ray computed maxillofacial region radiation exposure image quality
  • 相关文献

参考文献8

二级参考文献47

  • 1Pannu HK. Fishman EK. Multideteclur coinpuUMltdino^raphic evaluation of the renal arteryjj |.\l)(lom Imaging.2002.27(6):611-619.
  • 2Toprak O. Cirit M. Hisk factors and therapy strategies forcontrast-induced nr|)liro|>athy[J].Ken Kail, 2006, 28(5):36.S-38I.
  • 3M, Aspflin P. I^ameire N. Contrast-inducediH'plmipathy: a clinical aiul t*vi(lt*nc<*-l>astHl appn)arh[J].Circulation, 2006. 113(14):1799-1806.
  • 4INamimoto T, ()<ia S, Utsunomiya I), el al. Improvemenl ofimage (jualitv at 丨ow-radiati(m (lose a"(丨丨(>w-(‘onlmst materialdose alxiominal d,in patients with cirrhosis: intraiiulividualc(nnparis(.n of low tuht* voltage with ilt*rative reionstructionat^orillim ami slandard tube voltage]J).J (loinpul AssisiTom叩r,2012. 36(4):495-501.
  • 5Takeyanm N. Ohgiya Y, Hayaslii T, et al. Comparison ofdiflVrent volumes of saline flush in the assessment ofperivenous arlefads in tlie sulu-lavian vein <iuring cervicalCTangiography|J|.lirJ Kadiol, 2011, 84(1001):427-434.
  • 6iiae KP. FVak ronlrast enhancenifnt in CT and MKangiography: when does it occur and why. PliarniacHtkinelicstudy in a porcine mn(tel[J].r{a(liolo^. 2(K)3, 227(3):809-816.
  • 7Clieng J, Yin、. Wu II. et al. Oplimal 丨mmm.limmatk. energylevels ill speclra CT pulmonary angiography for tin*evaluation of piilmonan .,mholism|J|.PIx)S ()n.*. 2013. 8(5):e63140.
  • 8lie J. Ma X. W ang (^). et al. Spe<-tral (:T (ienionstration of thesuperior mesenlerif arlery: (-omparisoi) of immix-hroinatic andpolychnmiatic iniugin^;|JJ.Acai! Hadiol. 2014. 21(3):364-368.
  • 9G.d’Assignies,A.Couvelard,S.Bahrami,M.P.Vullierme,P.Hammel,O.Hentic,朱玉兆.胰腺内分泌肿瘤:灌注CT评价的肿瘤血流反映了血管发生并与预后因素相关[J].国际医学放射学杂志,2009,32(2):186-186. 被引量:57
  • 10张兆琪,徐磊.重视冠状动脉多层CT成像的低剂量检查[J].中华放射学杂志,2009,43(7):681-683. 被引量:80

共引文献74

同被引文献40

引证文献4

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部