PLUS移动X光机的工作原理及故障检修2例 被引量:1

Two Case Studies of Working Principle and Failure Maintenance of PLUS Mobile X-ray Machine
下载PDF
导出
摘要 本文对西门子POLYMOBIL PLUS移动式X光机的整机工作原理做简单介绍,对相应的电路进行了功能分析,提供相应的检修程序代码,并对使用过程中出现的两例特殊故障涉及电源供电故障和高压放电回路故障,进行详细地故障分析与检修。 In this paper,the structure and principle of Siemens POLYMOBIL PLUS Mobile X-ray machine was introduced.The function of the corresponding circuit was analyzed and provides the corresponding maintenance program code.The author provided analysis and judgment of two special failures involving machine power supply and high voltage discharge circuit failure appeared during use.The detailed fault analysis and maintenance process were introduced.
作者 杜嘉莉 DU Jiali(Department of Medical Equipment,The Third Hospital of Jinan,Jinan Shandong 250101,China)
出处 《中国医疗设备》 2021年第3期181-182,186,共3页 China Medical Devices
关键词 POLYMOBIL PLUS 移动X光机 工作原理 故障分析 POLYMOBIL PLUS mobile X-ray machine working principle troubleshooting